节点文献

婴儿髋关节超声检查床临床应用研究

Research on clinical application of the ultrasonic examining bed in infant hip joint

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 邓肖群杨晓娟管清宋焱峰崔立刚邓文怀

【Author】 DENG Xiao-qun;YANG Xiao-juan;GUAN Qing;SONG Yan-feng;CUI Li-Gang;DENG Wen-huai;Department of Ultrasound,The First People’s Hospital of Lanzhou;Department of Basic Medical Sciences,Lanzhou University;Department of Ultrasound,The Third Hospital of Peking University;Department of Materials Engineering,Lanzhou University of Technology;

【机构】 甘肃省兰州市第一人民医院功能科兰州大学基础医学院北京大学第三医院超声科兰州理工大学材料工程系

【摘要】 目的:探讨自制婴儿髋关节超声检查床在婴儿发育性髋关节发育不良超声筛查中的应用效果。方法:选择160名(320个髋关节)出生后42 d的婴儿进行髋关节超声检查。3名超声医生首先用普通检查床,徒手固定婴儿臀部及下肢进行检查;随后应用婴儿髋关节超声检查床再次检查,并对比在2种不同检查床条件下超声测量数据。结果:应用婴儿髋关节检查床总体检查的结果左右侧α角和β角差异均显著小于普通检查床检查的结果(P<0.01),右侧股骨头覆盖率检查的总体结果差异显著低于普通检查床检查的结果(P<0.01);使用普通检查床3名医生测量结果之间内部相关系数均<0.4,一致性较差,使用本研究设计婴儿髋关节超声检查床后内部相关系数均>0.75,一致性较好。结论:婴儿髋关节超声检查床的应用可明显减少人为因素误差,降低数据变异度。

【Abstract】 Objective: To investigate the effects of self-made ultrasonic examining bed in the screening of development dislocation of hip( DDH). Methods: Three hundred and twenty hip joints in 160 infants born more than 42 days were detected by ultrasound. The cases were examined using common examining bed combined with fixing the infant hip and lower limbs by hand by three doctors,and the cases were examined using infant hip joint ultrasonic examining bed again. The examining date between two methods were compared. Results: The deviation of α angle and β angle in left and right sides detected by infant hip joint examining bed were significantly less than those detected by the common bed( P < 0. 01). The difference of right side total result detected by FHC was significantly less than that of common bed( P < 0. 01). The intraclass correlation coefficient( ICC) of examination result using common bed by three doctors was less than 0. 4,the consistency of which was poor. The intraclass correlation coefficient of examination result using infant hip joint examining bed was more than 0. 75,the consistency of which was good. Conclusions: Ultrasonic examining bed for infant hip joint can significantly eliminate the anthropic factor influencing,and reduce the degree variation of date.

【基金】 甘肃省兰州市卫生科技发展资金项目
  • 【文献出处】 蚌埠医学院学报 ,Journal of Bengbu Medical College , 编辑部邮箱 ,2017年08期
  • 【分类号】R445.1;R726.8
  • 【被引频次】1
  • 【下载频次】50