瓦秒法检查单相电能表存在的误区及对策
The Existing Misunderstandings in Watt-second Method for Single-phase Electric Meter Checking and Its Countermeasures
【摘要】 针对基层供电公司采用瓦秒法现场检验单相电能表运行状况存在自带负载功率过小、接线错误及误解瓦秒法的检测结果含义等问题。本文在介绍瓦秒法原理的基础上,根据《电能计量装置技术管理规程》提出了自带负载的电流必须大于电能表标定电流的5%的建议;分析了自带负载各种接线方式的实质,得出自带负载必须接在电能表出线侧的结论;指出瓦秒法检测结果只反映电能表本身的运行误差,而不能反映电能表的接线误差。
【Abstract】 Aiming at the problems of Watt-second method is used in checking the condition ofa single-phase electric meter on spotby the grass-rootscompanies of power supply.It’s shown below:the self-taking power load used is toothin,its wiring mode is wrong,ortends to a mistakencon clusion from the checking results of Watt-second method,and soon.This paper firstly in troduced princi ples of Watt-second method,then proposed the suggestions that the current of self-takingload must be 5% greater than the calibrating current of the meter based on the technical management rules for electric energy metering device.Thirdly,the paper came toa conclusion that self-takingload must be linked tothe outlet side ofthe meter,after analyzing the substances ofvarious kindsofself-taking-load wiring modes,and lastly,pointed out that the results from Watt-second method only reflect the error from the meter,rather than from wiring modes.
【Key words】 Watt-second method; self-taking power load; wiring modes; error from energy meter; error from wiring modes;
- 【文献出处】 电测与仪表 ,Electrical Measurement & Instrumentation , 编辑部邮箱 ,2010年S2期
- 【分类号】TM933.4
- 【下载频次】63