节点文献
应用反相高效液相色谱法分离L,D-多肽非对映异构体
Resolution of L,D-Peptide Diastereoisomers with ReversedPhase High Performance Liquid Chromatography
【摘要】 应用反相高效液相色谱法(RP-HPLC)研究了5对L,D-多肽非对映异构体在不同条件下的分离情况。通过改换色谱柱、改变流动相的组成及梯度淋洗条件,发现其中的2对(Boc-Ala-L,D-Ala-GlyOBzl和Boc-Gly-L,D-Ala-ValOBzl)可以得到较好的分离,分离度分别为1.1和1.6,因此这2对L,D-多肽非对映异构体可以作为研究接肽反应消旋情况的模型体系。
【Abstract】 he resolution of 5 pairs of L,D - peptide diastereoisomers( Boc-Phe-GlyOBzl and Boc - D - Phe -GlyOBzl; Boc -Ala -Ala-GlyOBzl and Boc-Ala-D-Ala-GlyOBzl Boc-Ala-Phe-GlyOBzlandBoc-Ala-D-Phe-GlyOBzl; Boc-Ala-Leu-GlyOBzl and Boc-Ala-D-Leu-GlyOBzl ;Boc-Gly-L, D-Ala-ValOBzl) were investigated with reversed-phase high performance liquid chromatography under different conditions. The columns: NovaPak C18 (150×3. gmm i. d.), μ Bondapak C18 (250 × 3.9mm i.d.)μ Bondapak CN(150×3. 9mm i. d.),μBondapak Phenyl (300 × 3. 9mm i. d.),Synchropak RP-P (250×4. 6mm i. d.).Mobile phases:(1) A:0.1 % TFA-H2O, B: 0. 1% TFA-MeCNI (2) A: 0. 1% TPA-H2O, B: 0.1% TFA-MeOH. Detector wave length: 214nm.The optimum conditions obtained are Synchropak RP-P column with 20%-60% (1)B as gradient eluent for Boc-Ala-L, D-Ala-GlyOBzl and Novapak C18column with 30% (1 )B for Boc-Gly-L,D-Ala-ValOBzl. They were successively recognized into their chiralities respectively with 1. 1 and 1. 6 of resolution factors. It is recommended that the method can be applied to the racemzition study in peptide synthesis process.
【Key words】 high performance liquid chromatography; L; D-peptide; separation of diastereoisomers;
- 【文献出处】 色谱 ,CHINESE JOURNAL OF CHROMATOGRAPHY , 编辑部邮箱 ,1996年03期
- 【分类号】Q516
- 【被引频次】2
- 【下载频次】277