节点文献

液体闪烁测量125Iγ射线方法的研究

The measurement of gamma radiation of 125iodine by liquid scintillation counter

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 李迪元; 张昀; 陈贤妙; 曹素玲; 徐宝钢; 江浩;

【Author】 Li Diyuan Zhang Yun Chen Xianmiao Cao Suling Xu Baogang Jiang Hao(Shanghai Hypertension Research Institute)

【机构】 上海市高血压研究所; 上海市高血压研究所;

【摘要】 <正> 在液体闪烁液中加入重金属元素,就可用液体闪烁法测量γ射线。本文对125I的液体闪烁测定条件进行了探讨,并应用125I放射免疫的样本同时在γ谱仪和液体闪烁仪上测定并比较,确定125I液体闪烁测定的可行性。这样可以提高125I样本测定的自动化程度和实现液体闪烁仪一机多用。

【Abstract】 Tetrabutyl tin-loaded liquid scintillater has been used for gamma counting of 125iodine with a scintillation counter instead of the use of traditional gamma counter. The results of liquid scintillation counting measurements were closely agreed with those obtained by gamma counting method in the radioimmunoassays of some samples including PRA, Angiotension Ⅱ, enkphrine and bradikinin. The results show that this method may extend the liquid scintillation, to gamma counting.

  • 【文献出处】 核技术 ,Nuclear Techniques , 编辑部邮箱 ,1983年05期
  • 【下载频次】29
节点文献中: