节点文献
Luminol-H2O2化学发光体系测定水中对叔丁基苯酚
Flow Injection Chemiluminescence Determination of p-tert-butylphenol with Luminol-H2O2
【摘要】 采用流动注射技术研究了在碱性条件下,Luminol-H2O2与对叔丁基苯酚的化学发光行为,对影响化学发光强度的诸因素进行了优化,建立了流动注射化学发光法测定对叔丁基酚的新方法。方法的检出限为4×10-8g/L,线性范围为2.0×10-6~1×10-1g/L,对10-5g/L对叔丁基进行11次平行测定,相对标准偏差为1.45%。该法用于测定废水中对叔丁基苯酚的含量,结果令人满意。
【Abstract】 The chemiluminesecence behavior of the oxidation reaction of p-tert-butylphenol by Luminol-H2O2 was studied under the condition of flow-injection.The detection limit obtained was 4x10-8g/L.The linear range was 2.0×10-6g/L~1×10-4g/L.The precision was tested at the concentration level of 10-5g/L p-tert-butylphenol for 11 determinations and the value of the relative standard deviation obtained was 1.45%.This proposed method is satisfactory in the determination of p-tert-butylphenol in waste water.
- 【文献出处】 净水技术 ,Water Purification Technology , 编辑部邮箱 ,2006年03期
- 【分类号】X832
- 【被引频次】10