管电压对胸部DR影像质量的影响
Effects of the tube voltage on the chest image quality of digital radiography
【摘要】 目的研究有效剂量相等条件下不同管电压对胸部DR图像质量的影响。方法48例病人在90kV、125kV、150kV分别摄取后前位胸片,未经任何处理打印片子,五位资深放射专家采用五分法独立打分评估影像质量;采用Monte-Carlo转换因子估算有效剂量ED(mSv)。结果绝大多数解剖结构在90千伏时得分最高(平均得分3.10),除了在心影轮廓、气管隆突和外围纹理方面以外,其余部位均有明显统计学差异。结论有效剂量相等条件下采用较低千伏能获得更好胸部DR影像。
【Abstract】 Purpose To study the influence of different kilovoltages upon DR image quality of chest with effective dose eˉquivalence.Methods Posteroanterior chest radiographs of48patients were obstained at90,125,150kv,printed the raˉdiographs with out treatment,then five radiologists subjectively and independently rated on hard-copy images by using a five-point scale.The effective dose(ED)was estimated by Monte-Carlo conversion coefficients.Results The visiˉbility scores of most anatomic structures were significantly superior with the90kv(mean scroe of,3.10).The differences in rating scores did not reach statistical significance only for the peripheral vessels,the cardiac contours and the carina.Conclusion Most anatomic structures in chest were ranked superior in digital radiograph acquired with lower kilovoltage at a constant effective patient dose.
【Key words】 image quality; tube voltage; digital radiography; effective dose;
- 【文献出处】 医疗设备信息 ,Information of Medical Equipment , 编辑部邮箱 ,2006年06期