节点文献

流动注射测定水中氨氮条件的优化

The Optimization of Ammonium-nitrogen Determination Condition by flow Injection Analysis

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 张汪霞张新申缪培凯袁东

【Author】 ZHANG Wang-xia, ZHANG Xin-shen ,MJAO Pei-kai, YUAN Dong (Key laboratory of Leather Chemistry and Engineering (Sichuan university) , Ministry of Education, Chengdu 610065, China )

【机构】 皮革化学与工程教育部重点实验室(四川大学)皮革化学与工程教育部重点实验室(四川大学) 四川 成都 610065四川 成都 610065

【摘要】 采用流动注射次溴酸法测定江河水和海水中的氨氮,将其实验条件进行优化,改变氧化液的用量、进样体积等条件,可以更好提高测定的灵敏度和精确度。对方法的线性范围为5-400μg/L,检出限是2 μg/L,相对标准偏差(连续进样11次)是0.36%。

【Abstract】 Ammonium-nitrogen in seawater and estuarine water can be determined in flow injection analysis by ammonium reacts with hypo-bromate. Experiment conditions, the dosage of the oxidation solution loop volume and so on were optimized. The sensitivity and proecise of the method were improved a lot. The linear range in this condition was 5 g/L to 400 g/L. The detection limit was 2 g/L and the relative standard deviation (slope) is 0. 36% (n=11).

【基金】 国家863标准化工程课题(200410)-营养盐自动分析仪标准化工程
  • 【文献出处】 皮革科学与工程 ,Leather Science and Engineering , 编辑部邮箱 ,2006年03期
  • 【分类号】X832
  • 【被引频次】11
  • 【下载频次】163
节点文献中: 

本文链接的文献网络图示:

本文的引文网络