节点文献

聚苯胺修饰玻炭电极差示脉冲伏安法测定人体尿酸

DETERMINATION OF URIC ACID BY DIFFERENTIAL PULSE VOLTAMMETRY WITH GLASSY CARBON ELECTRODE MODIFIED BY POLYANILINE

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 黄桂萍钟平

【Author】 Huang Guiping, Zhong Ping (Department of Chemistry and Life Science Gannan Teachers’s College, Ganzhou 341000, China)

【机构】 赣南师范学院化学与生命科学系赣南师范学院化学与生命科学系 赣州341000赣州341000

【摘要】 研究了聚苯胺修饰玻炭电极的电化学性质和在抗坏血酸等物质存在下测定人体尿酸的分析方法。在0.025mol/L磷酸盐缓冲液(pH=6.6)介质中,聚苯胺修饰玻炭电极对尿酸具有强烈的吸附活性。在2.0×10-6~8.0×10-6mol/L内,尿酸浓度c与峰电流Ip成线性关系,相关系数r=0.9922。浓度为3.0×10-3mol/L的抗坏血酸不干扰尿酸的测定。用该方法测定了人体尿样中的尿酸含量,测定结果的相对标准偏差为3.21%,回收率为94.1%~101.6%。

【Abstract】 The electro-chemical characters of glassy carbon electrode modified by polyaniline and analytical method for uric acid in the presence of ascorbic acid were studied. In the buffer solution with pH of 6.6, there were absorption activity of glassy carbon electrode to uric acid and an excellent linear relation between the peak current and concentration in the range of 2.0×10 -6 -8.0×10 -6 mol/L (r=0.9922). 3.0×10 -3 mol/L ascorbic acid didn’t interfere with the determination of 2.0×10 -6 mol/L uric acid. The RSD of determination results was 3.21% and recoveries were 94.1%-101.6%.

  • 【文献出处】 化学分析计量 ,Chemical Analysis and Meterage , 编辑部邮箱 ,2006年01期
  • 【分类号】O657.1
  • 【被引频次】7
  • 【下载频次】287
节点文献中: 

本文链接的文献网络图示:

本文的引文网络