节点文献
CTMAB增敏催化动力学光度法测定痕量锰
Determination of Trace Manganese by Catalytic Kinetic Spectrophotometry Method with CTMAB as the Sensitizer
【摘要】 研究了以NTA为活化剂,CTMAB增敏,Mn(Ⅱ)催化KIO4-氧化7-(2-硝基-4-甲基苯偶氮)-8-羟基喹啉-5-磺酸(简称7P8Q5S)的反应,建立了催化光度法测定痕量锰的新方法,以509 nm为测定波长,Mn(Ⅱ)在0~3.2×10-9g.mL-1的范围内成线性关系,检出限为3.8×10-11g.mL-1,考察了20多种共存离子的影响,用于测定大米、绿豆、小米和土壤中的锰。
【Abstract】 A new catalytic spectrophotometric method for the determination of trace amount of Mn(Ⅱ) was developed.The method is based on the oxidation of 7-(2-nitryl-4-methyl phenylazo)-8-hydroxy-qinolin-5-slfonic acid by potassium periodate in the presence of nitrilo triacec acid as an activator and CTMAB as a sensitizer.The wavelengths for the determination was 509 nm.The linear range of determination is 0-3.2×10-9 g·mL-1 for Mn(Ⅱ).The determination limit is 3.8×10-11 g·mL-1.The interference of foreign ions was studied.The method can be used for the determination of Mn(Ⅱ) in mung bean,rice,cereal and soil samples.
【Key words】 CTMAB; 7-(2-nitryl-4-methyl phenylazo)-8-hydroxy-qinolin-5-slfonic acid; Catalytic spectrophotometric method; Nitrilo triacec acid; Potassium periodate; Manganese;
- 【文献出处】 光谱学与光谱分析 ,Spectroscopy and Spectral Analysis , 编辑部邮箱 ,2006年08期
- 【分类号】O614.711;O657.3
- 【被引频次】9