心率对64排螺旋CT冠状动脉成像质量的影响及对策

Impact of Image Quality of Heart Rate on Coronary CT Angiography with 64-Detector Row CT and the Relevant Measurement of Solution

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 谢元亮郑晓华金朝林张树桐王俊陈发祥王翔

【Author】 XIE Yuan-liang,ZHENG Xiao-hua,ZHENG Xiao-hua,et al.The People′s Hospital of Wuhan University,Wuhan 430060,P.R.China

【机构】 武汉大学人民医院放射科武汉市中心医院影像科武汉市中心医院影像科 430060武汉430014武汉

【摘要】 目的:探讨心率对64排CT冠状动脉成像质量的影响,以及运用多时相重建、多点重建、心电图(ECG)编辑等技术的价值。方法:353例患者行64排螺旋CT冠状动脉成像,对75%R-R间期重建图像上存在伪影的202例,进行多时相重建和/或多点重建和/或ECG编辑再重建。根据心率及心率变化分组,评价不同心率、不同心率波动时和采用3种重建方法时的图像质量。结果:本组中心率≤65bpm组图像合格率为98.3%,66~70bpm组为94.2%,>70bpm组为69.2%。心率>70bpm组和/或心率变化大于10%时图像质量均显著下降(P<0.01)。59.3%(35/59)不可评价图像经多时相重建和/或多点重建和/或ECG编辑再重建后成为可评价图像。结论:64排螺旋CT冠状动脉成像时,心率低于70bpm且波动小于10%能获得较好的图像质量;心率>70bpm和/或波动>10%者,应用多时相重建、多点重建和ECG编辑再重建能明显提高图像质量。

【Abstract】 Objective:To investigate the impact of image quality of heart rate on coronary CT angiography (CTA),and the value of multi-phase,multi-spot as well as ECG editing reconstruction.Methods:Coronary CTA was performed in 353 patients with 64-detector row CT.In 202 patients,which had 75% of the images with artifacts after R wave,reconstruction with multi-phase or multi-spot of cardiac cycle or/and with ECG editing were carried out.According to the subgroups divided by heart rate and heart rate fluctuation,image quality of various heart rate,various situation with heart rate fluctuation and 3 different approaches of reconstruction were evaluated.Results:Qualified images with heart rate ≤65bpm,66~70bpm,>70bpm accounts for 98.3%,94.2% and 69.2% respectively.Marked downgrading of image quality were shown in the groups of heart rate >70bpm and /or heart rate fluctuation >10% (P<0.01).Of the 59 non-qualified images,35 (59.3%) were upgraded to be qualified after multi-phase,multi-spot and/or ECG edited reconstruction.Conclusion:Using 64-detector row CT,qualified coronary angiography images could be achieved in patients with heart rate <70bpm and the fluctuation <10%.Reconstruction with multi-phase or multi-spot of cardiac cycle or/and with ECG editing was helpful to improve the image quality.

  • 【文献出处】 放射学实践 ,Radiologic Practice , 编辑部邮箱 ,2006年09期