节点文献
西安脉冲堆中子照相系统性能测试及应用
Test and Application of Thermal Neutron Radiography Facility at Xi’an Pulsed Reactor
【摘要】 介绍了西安脉冲堆中子照相系统的组成以及对该系统进行性能测试的方法与结果。结果表明, 所用BAS-ND中子屏的固有不锐度为0.15mm;在稳态堆芯及脉冲堆芯布置下,中子照相孔道射线束中的有效热中子含量偏低, 而有效γ射线含量偏高;在孔道中采用铅来过滤射线束能明显降低有效γ射线含量。
【Abstract】 A thermal neutron radiography facility at Xi抋n Pulsed Reactor is described as well as its characteristics and application. The experiment results show the inherent unsharpness of BAS ND is 0.15mm. The efficient thermal neutron n/γ ratio is lower in not only steady state configuration but also pulsing state configuration and it is improved using Pb filter.
【关键词】 中子照相;
灵敏度指示器;
束纯度指示器;
中子屏;
脉冲反应堆;
【Key words】 Neutron radiography; Beam purity indicator; Sensitivity indicator; Neutron imaging plate; Pulsed reactor;
【Key words】 Neutron radiography; Beam purity indicator; Sensitivity indicator; Neutron imaging plate; Pulsed reactor;
- 【文献出处】 核动力工程 ,Nuclear Power Engineering , 编辑部邮箱 ,2002年06期
- 【分类号】TL411.4
- 【下载频次】81