节点文献

基于溶胶-凝胶技术可更新的安培分析免疫传感器用于人血清中补体3的测定

Renewable Amperometric Immunosensor Based on Sol-Gel Technique for Complement 3 Assay

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 钟桐生刘国东黄杉生沈国励俞汝勤

【Author】 ZHONG Tong sheng 1,2 , LIU Guo dong 1, HUANG Sha sheng 1* , SHEN Guo li 1, YU Ru qin 1 (1. College of Chemistry and Chemical Engineering, Institute for Chemometrics and Sensing Technology Hunan University, Changsha, 410082, P. R. Chi

【机构】 湖南大学化学化工学院湖南大学化学化工学院 长沙410082益阳师范高等专科学校湖南益阳413049长沙410082410082

【摘要】 在低温下 ,通过混合石墨粉、补体 3抗血清、溶胶 -凝胶而制备了一种基于溶胶 -凝胶技术的可更新的安培分析免疫传感器。该传感器坚固、多孔 ,拥有一个可更新的表面 ,在辣根过氧化物酶标记的补体 3的协助下 ,通过竞争性的免疫分析来确定人血清中补体 3的含量。以邻 -氨基酚作为底物 ,- 1 5 0 m V作为工作电位 ,响应电流与补体 3的浓度在 1 .1 7~ 35 .1 μg/m L范围内呈线性关系 ,检出限为 0 .5 6μg/m L。分析血清样本结果表明所制备的传感器在临床分析中是可行的

【Abstract】 A renewable amperometric immunosensor based on sol gel technique has been constructed by dispersing graphite, complement 3(C 3) antiserum and sol gel at low temperature. The prepared immunosensor is rigid, porous, and has a renewable external surface. A competitive binding assay has been used to determine C 3 in human serum with the aid of C 3 labeled with horseradish peroxidase. The enzyme labeled antigen can readily diffuse toward the encapsulated antibody, which retains its binding properties. The experimental parameters of the assay with the biocomposite, including the loading of C 3 antiserum in the biocomposite, the amount of labeled C 3 in incubation solution, incubation time and temperature, have been optimized. Using C 3 labeled with horseradish peroxidase, and o AP as the substrate, an amperometric detection at -150 mV (vs. SCE) results in a linear detection range of 1.17~35.1 μg/mL, with a detection limit of 0.56 μg/mL. Serum samples have been assayed and the results demonstrate the feasibility of the proposed immunosensor for clinical analysis. The surface of the immunosensor can be renewed by simple polishing to obtain a fresh immunocomposite ready to use in a new competitive assay.

【关键词】 可更新免疫传感器补体3
【Key words】 RenewableSol gelImmunosensorComplement 3
  • 【文献出处】 分析科学学报 ,Journal of Analytical Science , 编辑部邮箱 ,2002年01期
  • 【分类号】O657.1
  • 【被引频次】9
  • 【下载频次】62