节点文献

经Swartz鞘标测和消融常规方法消融失败的房室旁路

Catheter Mapping and Ablation of Unsuccessful Atrioventricular Accessory Pathway Ablated by Conventional Methods Through Swartz Sheath

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 王业松马虹唐安丽柳俊吴素华何建桂廖新学胡苑

【Author】 WANG Ye-song, MA Hong, TANG An-li, LIU Jun, WU Su-hua, HE Jian-gui, LIAO Xin-xue, HU Yuan (Department of Cardiology, First Affiliated Hospital, Sun Yat-sen University of Medical Sciences, Guangzhou 510080, China)

【机构】 中山医科大学附属第一医院心内科!广东广州510080中山医科大学附属第一医院?

【摘要】 【目的】评价Swartz鞘在房室旁路标测和消融中的作用。【方法】 2 3例常规导管标测和消融未成功的房室旁路 ,采用消融导管经Swartz鞘标测和消融的方法。【结果】 2 3条旁路全部消融成功。 11条右外侧显性旁路用SR2 型和SR3 型Swartz鞘、12条左前外侧旁路 (显性 10条、隐匿性 2条 )用SL1和SL2 型Swartz鞘。标测和消融的操作时间 (4 6± 2 0 )min ,X线曝光时间 (2 4± 12 )min。右外侧和左前外侧显性旁路成功消融靶点图的比较 :A/V比值为 0 6 2± 0 32比 0 6 5± 0 2 8(P >0 0 5 )、AV间期为 (2 9± 12 )ms比 (37± 11)ms(P <0 0 5 )、VΔ间期为 (18± 11)ms比 (8± 9)ms(P <0 0 5 )、靶点图的稳定性为 0 74±0 16比 0 78± 0 12 (P >0 0 5 )。随访 (16± 14)月 ,无 1例复发。【结论】Swartz鞘有助于消融导管标测定位和消融导管头端电极与房室环紧密接触 ,从而提高消融成功率 ,减少复发 ,缩短操作时间。

【Abstract】 Objective To evaluate the effect of Swartz sheath on mapping and ablation of atrioventricular accessory pathway(AP). Methods 23 cases of atrioventricular AP failed by conventional catheter mapping and ablation techniques. They were mapped and ablated by catheter inserted through Swartz sheath. Results 23 atrioventricular AP in the 23 cases were successfully ablated. SR 2-3 Swartz sheaths were utilized in 11 right lateral overt AP. SL 1-2 Swartz sheaths were utilized in 12 left anterolateral AP (10 overt AP and 2 concealed AP). The mean procedure time of mapping and ablation was (46±20) min and the mean fluoroscopy time was (24±12) min. The right lateral overt AP were compared with the left overt AP in local endocardial electrograms recorded at successful ablation sites, which showed that A/V ratio was 0.62±0.32 vs 0.65±0.28(P>0.05), AV interval was (29±12) ms vs (37±11) ms(P<0.05), V-dalta interval was (18±11) ms vs (8±9) ms(P<0.05), and stability of atrial and ventricular electrograms was 0.74±0.16 vs 0.78±0.12(P>0.05). During a mean follow-up of (16±14) months,there is no recurrent case. Conclusion Swartz sheath not only contributes to catheter mapping and AP localization but also close contact of catheter tip electrode with the ring of atrioventricular valve. So the successful rate of ablation was improved,and the recurrent rate and procedure time were reduced.

  • 【文献出处】 中山医科大学学报 ,Academic Journal of Sun Yat-sen University of Medical Sciences , 编辑部邮箱 ,2001年04期
  • 【分类号】R541.7
  • 【被引频次】3
  • 【下载频次】33
节点文献中: 

本文链接的文献网络图示:

本文的引文网络