节点文献

冠状动脉内置入国产支架的动物实验研究

The experimental study on intracoronary native metallic balloon-expandable stent

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 朱锐江洪李庚山漆曙辉王小红王洪如叶泽兵

【Author】 Zhu Rui, Jiang Hong, Li Gengshan, et al. (Department of Cardiology, First Affiliated Hospital of Hu Bei Medical University, Wuhan 430060)

【机构】 湖北医科大学附属第一医院心内科!湖北省襄樊市中心医院心内科武汉430060湖北医科大学附属第一医院心内科!武汉430060湖北医科大学附属第?

【摘要】 目的观察国产支架置入大冠状动脉后的生物、血液相容性、支架开通率及其内皮化情况。方法将16个国产支架置入8只犬的冠状动脉内,分别于1、2、3、6个月进行观察。结果支架置入术后冠脉通畅,无明显狭窄。支架贴内膜良好,血管内腔表面光滑。术后2个月时支架覆盖的新生内膜最厚,其主要为平滑肌细胞和细胞外基质,未见明显炎性细胞浸润。支架置入1个月时,支架表面覆盖的内皮细胞排列不规则;置入2、3、6个月时,其沿血流方向排列规则。支架表面未见明显腐蚀。结论该国产支架置入后有良好的生物和血液相容性,理化性能稳定,迅速重新内皮化,有良好的开通率。

【Abstract】 Objective To investigate the biocompatibility, bloodcompatibility, the patency rates and reendothelialization of native coronary metallic stent after being implanted into canine coronary artery.Methods Sixteen stents were implanted into eight canine coronary artery through femoral artery aproach.The pathological examination was taken at 1 to 6 months after coronary angiography.Results Thirteen stents were successfully implanted into target vessels of six dogs. Two dogs died of coronary spasm and acute stent thrombosis respectively. Coronary angiography demonstrated all vessles patency after intracoronary stenting (ICS). Macroscopy of stented arteries demonstrated complete incorporation of stent wires into the arterial wall and smooth vessle lumens. Light microscope demonstrated the neointima was the thickest at two months and consisted mainly of smooth muscle cells and extracellular matrix. The findings also included slight to moderate media oppression localized to areas of stent wires and no significant inflammatory raction. Scanning electron mcroscope demonstrated the endotheliocytes covering on stents oriented irregularly at one month and regularly at two and three as well as six months in the direction of blood flow. There was no significant erosion on stent wires.Conclusion The native balloon- expandable coronary metallic stems are promisint with good biocompatibility, biocompatibility and physicochemical stability after being implanted into canine coronary arteries. They can maintain high patency rates due to rapid reendothelialization.

【基金】 湖北省“九五”攻关项目资助
  • 【文献出处】 中国介入心脏病学杂志 ,Chinese Journal of Interventional Cardiology , 编辑部邮箱 ,2001年01期
  • 【分类号】R-332
  • 【被引频次】4
  • 【下载频次】66
节点文献中: 

本文链接的文献网络图示:

本文的引文网络