节点文献

差热分析法测定化工品、药品纯度

PURITY DETERMINATION OF CHEMICALS & PHARMACEUTICALS BY DSC METHOD

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 纪雷王岩王英杰徐枫

【Author】 Ji Lei, Wang Yan, Wang Yingjie, Xu Feng (Qingdao Entry - Exit Inspection and Quarantine Bureau 266002)

【机构】 青岛出入境检验检疫局青岛出入境检验检疫局 266002266002266002

【摘要】 研究了差热分析(DSC)法测定化工品、药品纯度的条件,测定的最佳条件为升温速率0.7℃/min,样品量2~3mg,保护气流速20~40mL/min。测定结果的RSD为0.02%,F检验及t检验的结果表明,该方法与容量法无显著差异,该方法简便、快速、准确,适合化工品、药品的定量测定。

【Abstract】 The purity determination condition of chemicals and pharmaceuticals by DSC technique was introduced. The optimal testing conditions are: heating ramp being 0.7℃/min, sample weight being 2 - 3 mg and gas flow rate being 20 - 40 mL/min. The experiments yielded good result: there is no significance between DSC and conventional method by F-test and t-test, and RSD is 0.02% . It can be used in the quantative determination of chemicals and pharmaceuticals with its accuracy, convenience and practicality.

  • 【文献出处】 化学分析计量 ,Chemical Analysis and Meterage , 编辑部邮箱 ,2001年06期
  • 【分类号】O657.9
  • 【被引频次】17