节点文献

A/B超交替测量眼轴用于人工晶体度数计算

Axial Eye length measurement with alternative A/B scan for IOL power calculation

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 顾真寰; 沈慧英; 盛敏杰;

【Author】 Zhenhuan Gu, Hidying Shen, Minjie Sheng, Shanghai Railway Hospital, Shanghai 200072, China

【机构】 中国上海同济大学医学院附属铁路局中心医院眼科; 中国上海同济大学医学院附属铁路局中心医院眼科 200072; 200072;

【摘要】 目的 探索并发性白内障玻璃体混浊时准确测量眼轴的新方法。方法 用A超测量角膜到晶体后囊的距离,利用B超直观性测量晶体后囊到视网膜的距离,两者相加得出眼轴长度用于人工晶体计算,可避免将玻璃体混浊误认为眼后壁。结果 与常规A超测量法相比,术后残余屈光在1D以上的比例减少12%。结论 本测量法设计合理,准确性高,实用性强。

【Abstract】 Objective To explore new means to measure axial eye length intercurrent cataract and inpatients with vitreous turbidity . MethodsThe distance between the corneal and posterior capsular. was measured by A scan, and the distance between the posterior capsular and retinalwas measured by was B scan measured.The avial eye length was reached by adding these two results and was available for IOL calculation.Results Compared with routine A scan measurement, the cases with postoperative unexpected refractive error of 1 dioptre or above werereduced by 12% .Conclusion The new measurement is reasonable,accurate and practicabe.

  • 【分类号】R779.66
  • 【被引频次】4
  • 【下载频次】75