节点文献

ICP-MS法直接测定冰芯样品中超痕量镉

Direct determination of Cd in ice core samples by ICP-MS

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 李月芳; 唐富荣;

【Author】 LI Yue fang and TANG Fu rong (Ice Core and Cold Region Envionment Research Laboratry, Cold and Arid Regions Environment and Engineering Research Institute , Chinese Academy of Science,730000)

【机构】 中国科学院寒区旱区环境与工程研究所冰芯与寒区环境研究开放实验室!兰州730000;

【摘要】 采用 ICP- MS法对冰芯样品中超痕量 Cd的直接测定进行了研究。确定了直接测定浓度为 pg/m L 级的 Cd的最佳仪器参数、载气流速、进样速度等与灵敏度之间的关系以及浓度和扫描参数对分析精度的影响。本方法对浓度 5~2 0 0 pg/m L Cd的分析 ,RSD<1 0 % ,回收率在 88%~ 1 0 5%之间 ,检测限为0 .1 5pg/m L

【Abstract】 A new method for determination of trace Cd in ice core samples by ICP MS was developed. The analysis parameters for measuring Cd was optimized. The relationship of sensitivity with flow rate of sample gas and auxilliary gas were studied. The influence of concentrations and scan parameters on analysis precision were also studied. The precision(RSD) is less 10%, corresponding to the concentration range of 5~200 pg/mL. The recovery from standard solutions is in the range of 88%~105%. The detection limit is 0.15 pg/mL.

【关键词】 ICP-MS; 镉; 冰芯;
【Key words】 Trace Cd; Ice core samples; ICP MS;
【基金】 国家自然科学基金 !(4 980 1 0 0 4 );国家重点基础研究发展规划 !(G1 9980 4 0 80 0 );中国科学院知识创新工程项目!(KZCX1 -1 0 -0
  • 【文献出处】 分析试验室 ,Analytical Laboratory , 编辑部邮箱 ,2001年03期
  • 【分类号】O657.63
  • 【被引频次】9
  • 【下载频次】102
节点文献中: 

本文链接的文献网络图示:

本文的引文网络