非平衡流动注射分光光度法在线测定痕量铅的研究

Non-equilibrium Flow Injection Spectrophotometry for Pb2+ On-line Determination

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 訾言勤陈立国陈运生

【Author】 ZI Yan-qin, CHEN Li-guo, CHEN Yun-sheng ( Department of Chemistry, Huaibei Coal Teachers College, Huaibei 235000, China; Huaibei Environment monitoring station, Huaibei 235000, China)

【机构】 淮北煤炭师范学院化学系!安徽淮北235000淮北市环境监测站!安徽淮北235000

【摘要】 采用流动注射分析技术,研究了在碱性介质中铅与meso-四-(4-三甲钱基苯)卟琳高灵敏的显色反应,用自制的微机化流动注射分析仪准确控制反应时间,利用非平衡法原理,优化了实验条件,建立了在线测定痕量铅的新方法。线性范围为0.00~2.00 mg/L,分析速度每小时测定120次。方法适用于直接测定煤矿井水、鱼塘水、电场废水和自来水中的铅,所得的结果令人满意。

【Abstract】 The colour reaction of Pb2+ with meso -tetrakis -(4- trimethylammoniumphenyl ) porphyrin in alkaline medium was studied. The experimental conditions were optimized by using self- made micro- computerized flow injection spectrophotometric analyzer, and a new analytical method for the on -line determination of Pb2+ based on non - equilibrium principle was established. The linear range of the method is 0.00-2.00 mg/L for Pb2+, and its analytical rate is 120/h. It is suitable for the determi- nation of trace lead in mine water, fishpond water, power plant waste water and well water with satis- factory results.

【基金】 安徽省工业重点基金资助项目(皖科字92-19H8);煤炭工业部科研基金资助项目(煤科字96-374)
  • 【文献出处】 分析测试学报 ,Journal of Instrumental Analysis , 编辑部邮箱 ,2000年06期
  • 【分类号】O657.3
  • 【被引频次】12
  • 【下载频次】65