节点文献
化学发光测定TSH的实验评价与临床应用(英文)
EVALUATION AND CLINICAL PRACTICE OF CHEMILUMINESCENT IMMUNOASSAY DETECT THYROTROPIN
【摘要】 目的 :评估化学发光测定TSH的实验性能 ,拓宽TSH在临床诊断上的应用。方法 :采用“第三代”化学发光免疫分析“夹心法”。结果 :该法精密度试验批内与批间CV分别为 2 0 8- 2 74%与 3 85 - 4 43 % ,回收率10 0 6 % ,灵敏度 0 0 1mIU/L ,交叉污染率 <0 0 1% ,与RIA法对比试验Y =0 8942X + 0 2 0 31(r=0 986 7)。结论 :化学发光法能提供高灵敏度与高性能的测定TSH方法 ,同时通过TSH测定不但能区分甲状腺功能减退而且能应用于诊断甲状腺功能亢进 ,甚至亚临床型或潜在型甲状腺功能亢进或减退。
【Abstract】 Objective: To establish the chemiluminescent immunoassays for edtection of thyroid-stimulating hormone and enlarge upon clinical diagnostic value. Methods: Using highly sensitive "third generation" chemiluminescent immunoenzymatic "Sandwich" assay. Results: The intraassay and interassay imprecision were 2.08-2.74% and 3.825-4.43% The average rate of spiking recovery was 100.6%. The functional sensitivity was 0.01 mIU/L. The carryover was found to be<0.01%. This assay showed good linearity across the analytical radioimmunoassay Y=0.8942X+0.203(r=0.9867). Conclusion: This assay can offer highly sensitive quantitative detect thyrotropin with which to distinguish hypothyroidism from nonthyroideal disease and applied to diagnosis hyerthyroidism even subclinical and latent hyerthyroidism or hypothyroidism.
【Key words】 Chemiluminescent immunoassay (CLIA); Thyrotropin (TSH); Hyerthyroidism; Hypothyroidism;
- 【文献出处】 泸州医学院学报 ,JOURNAL OF LUZHOU MEDICAL COLLEGE , 编辑部邮箱 ,2000年05期
- 【分类号】R446.1
- 【下载频次】27