应用4650Ⅰ型压缩机雾化器进行支气管激发试验的研究

Application of Model 4650 Type Ⅰ Compressor Atomizer in Bronchial Challenge Test

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 袁玉如王佑娟曾继军何太灵

【Author】 Yuan Yuru; Wang Youjuan; Zeng Jijun; He Tailing (Department of Internal Medicine, The First Affiliated Hospital, WCUMS,Chengdu 610041)

【机构】 华西医科大学附属第一医院内科!成都610041

【摘要】 为了解Models4650Ⅰ型压缩机雾化器(简称M)进行支气管激发试验的可行性,将92例支气管哮喘息者按用M雾化吸入时间的不同,分为1、1.5、2分钟三级,并与标准雾化装置Dosimeter及雾化器(简称D)进行支气管激发试验同步对比研究。结果显示,三组受试者用两种雾化装置所获测值间无显著性差异(P>0.2~0.5),且皆呈直线正相关。M与D皆雾化吸入1分钟时,M所需药物浓度大于D,相关系数(0.285)偏低,延长M的雾化吸入时间至1.5分钟,与D雾化吸入1分钟相比,两者差值最小(-0.075g/L),r=0.665;若再延长M雾化吸入时间,与D的差值又会扩大。提示:M雾化效果稍差于D,适当延长时间可弥补上述差异,建议M雾化吸入时间以1.5分钟为宜。

【Abstract】 This study was directed to the feasibility of applying a simple atomizer-model 4650 type Ⅰ (abbrev. M)-to bronchial challenge test. 92 cases of asthma were divided into 3 groups randomly. All of them were subjected to the bronchial challenge test by M atomizer, and by standard Dosimeter atomizer (abbrev. D) as a comparison. In the test by M atomizer, the times for inspiring challenging medicine were 1, 1. 5 and 2 minutes for the 3 groups respectively, while the time for D atomizer was l minute for all. The results showed no significant differences (P>0. 2-0. 5) between the two atomizers in the 3 groups, their values were linear correlated. When the inspiring time was 1 minute for both M and D, the test needed a higher concentration of challenging medicine for M than for D, their coefficient of correlation (r=0. 285) was relatively low- When inspiring time postponed to 1. 5 minute for M, the difference in medicine concentration between M and D was smallest (- 0. 075g/L), r= 0. 665. However, a further postponed inspiring time to 2 minute for M reversely broadened their difference. These results indicated that the efficiency of M atomizer was a little lower than that of D, postponing the inspiring time for M could make up this weakness. A 1. 5 minute inspiring time for M atomizer was the suggestion. Some modifications on M atomizer were done by us for a better efficiency, and the cheap and popular M atomizer could be a good replacement in bronchial challenge test.

  • 【文献出处】 华西医科大学学报 ,JOURNAL OF WEST CHINA UNIVERSITY OF MEDICAL SCIENCES , 编辑部邮箱 ,2000年02期
  • 【分类号】R446
  • 【下载频次】34
节点文献中: 

本文链接的文献网络图示:

本文的引文网络