节点文献
用改造后的柯氏干涉仪测量量块中心长度的测量不确定度评定
Uncertainty Evaluation of the Measurement of Gauge Block by Using an Improved Koester Interferometer
【摘要】 通过用633nm 短频He-Ne 激光代替氪灯,提高小数测量的准确度等措施,对常规柯氏干涉仪进行了改造,结果获得测量量块中心长度的标准不确定度为uc= (5.5nm )2+ (8.3×10- 8L)2,大大优于一等量块的最大允许不确定度
【Abstract】 By the improvements such as using a stabilized He\|Ne laser instead of Kr lamp,increasing the measurement uncertainty of the fraction order,etc.,the standard unceratinty of the measurement of the length of gauge block by using a Koester interferometer was obtained.It is much more better than the maximum permissible uncertainty of the 1st order gauge block.
- 【文献出处】 现代计量测试 ,MODERN MEASUREMENT AND TEST , 编辑部邮箱 ,1999年06期
- 【分类号】TH744.3
- 【被引频次】10
- 【下载频次】76