节点文献

用顶空气相色谱法测定西伐他汀中甲苯残留量

Determination of Residual Toluene in Simvastatin by Headspace Gas Chromatography

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 李月微霍燕兰

【Author】 Li Yuewei, Huo Yanlan (Guangzhou Pharmaceutical Secondary Technical School, Guangzhou 510430)

【机构】 广州市医药中等专业学校实验室

【摘要】 建立了用顶空GC法测定西伐他汀中甲苯残留量的方法,选用401有机担体作色谱柱,用50g/L十二烷基硫酸钠水溶液作溶剂。线性范围为0.5~4.0mg/L,r=0.9987;平均回收率为103.06%。该法灵敏度高、简便、准确。甲苯的最低检测限为0.5mg/L。

【Abstract】 A method for the determination of residual toluene in simvastatin by headspace gas chromatography with a GDX-401 column was established. A 50 g/L sodium dodecyl sulfate solution as solvent of sample was used. The result showed that the linear range of calibration curve was 0.5~4.0 mg/L (r=0.998 7); recovery was 103.06 %(n=3); the detection limit of toluene was 0.5 mg/L. The method is simple, accurate and sensitive.

  • 【文献出处】 分析测试学报 ,JOURNAL OF INSTRUMENTAL ANALYSIS , 编辑部邮箱 ,1999年01期
  • 【分类号】TQ460.72,O657.71
  • 【被引频次】6
  • 【下载频次】58
节点文献中: 

本文链接的文献网络图示:

本文的引文网络