节点文献

流动注射-化学发光法测定盐酸羟胺

FIA-CL Determination of Hydroxyammonium Chloride

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 吕继宏李保新宋正华

【Author】 LU Ji-hong,LI Bao-xing,SONG Zheng-hua (Instrumental Analysis Research Center,Northwest University,Shanxi Xi′an 710069,China)

【机构】 西北大学分析测试研究中心!陕西西安710069

【摘要】 基于盐酸羟胺在弱酸性介质中被KIO4 氧化产生H2 O2 , H2 O2 与Luminol 发生化学发光反应,设计了流动注射—化学发光测定盐酸羟胺的方法. 线性范围为0-03 ~1-0μg/mL; 检出限为6-7ng/mL; RSD 为3-6 % . 测定含盐酸羟胺试样, 回收率为95 % ~98 % .

【Abstract】 A new method for determination of hydroxyammonium chloride with FIA-CL was developed in this paper.The principle of the method is that the hydroxyammonium chloride reacted with KIO 4 to generate H 2 O 2 in weak acid medium,and by chemiluminescence measurment of luminol- H 2 O 2 system,hydroxyammonium chloride was indirectly determined.The linear range was from 0.03~1 0μg/mL.RSD was 3 6% and the detection limit was 6 7ng/mL.A complete analysis can be performed within 2 minutes.The influence of foreign compounds was studied.The method was applied to determination of hydroxyammonium chloride in samples,and the recovery was from 95% to 98%.

  • 【文献出处】 福州大学学报(自然科学版) ,JOURNAL OF FUZHOU UNIVERSITY(NATURAL SCIENCES EDTION) , 编辑部邮箱 ,1999年S1期
  • 【分类号】O657.3
  • 【被引频次】4
  • 【下载频次】91
节点文献中: 

本文链接的文献网络图示:

本文的引文网络