液相色谱-串联质谱法测定晚期实体癌患者静脉滴注紫杉醇胶束后的药动学
A pharmacokinetic study of paclitaxel-loaded micelle in patients with advanced solidcarcinoma by LC–MS/MS
【作者】 茅晓寅; 张梦琪; 周沁逸; 董春霞; 沈薇薇; 刘罡一; 贾晶莹; 余琛; 史美祺;
【Author】 MAO Xiao-yin;ZHANG Meng-qi;ZHOU Qin-yi;DONG Chun-xia;SHEN Wei-wei;LIU Gang-yi;JIA Jing-ying;YU Chen;SHI Mei-qi;Central Laboratory, Shanghai Xuhui Central Hospital;Department of Internal Medicine, Jiangsu Cancer Hospital;
【机构】 上海市徐汇区中心医院中心实验室; 江苏省肿瘤医院内科;
【摘要】 目的建立测定人血浆中紫杉醇浓度的液相色谱-串联质谱(LC-MS/MS)方法。方法 8例晚期实体癌患者,进行单次静脉滴注注射用紫杉醇胶束(300mg·m-2),采集血浆样本并测定其中紫杉醇的浓度。血浆样本以氘5-紫杉醇为内标,血浆经直接沉淀后进样分析,选用CAPCELL PAK C18 MGIII(100 mm×2.0mm,5μm)为分析柱,以0.2%甲酸的水溶液-乙腈溶液=40∶60(V/V)为流动相,流速为0.4 mL·min-1。选用三重四级杆串联质谱仪的多重反应监测(MRM)扫描方式进行监测,电喷雾离子化源,正离子方式。并采用Phoenix WinNonlin 6.2软件对数据进行处理,计算药动学参数。结果血液中紫杉醇的线性范围10~20000μg·L-1。日内、日间RSD均小于8%,平均提取回收率在89.5%~97.7%范围内。内标校正基质因子为0.888 7~1.033,RSD<4%。应用此法,对8例晚期实体癌患者静脉滴注注射用紫杉醇胶束(300 mg·m-2)所得主要药动学参数为:ρmax为(3 872±1 062)μg·L-1,AUC0-∞为(14603±3390)μg·h·L-1,t1/2为(18.3±6.4)h,CL为(22.0±5.0)L·h-1·m-2。结论建立的LC-MS/MS测定人血浆中紫杉醇浓度的方法,灵敏度高、专一性好、操作简单,可用于临床药动学的研究。
【Abstract】 AIM To establish a liquid chromatography-tandem mass spectrometric(LC-MS/MS) method for determination of paclitaxel in human plasma. METHODS Plasma samples were collected from 8 patients with advanced solid cancer after administered intravenous paclitaxel-loaded micelle infusion at a single dose of 300 mg·m-2. In order to determine the concentration of paclitaxel, plasma samples were spiked with D5-labeled paclitaxel(internal standard,IS) and precipitated by methanol to remove protein before analysis. Chromatographic separation was achieved by a CAPCELL Pak C18 MGIII(100 mm×2.0 mm, 5μm) with an isocratic mobile phase which consisted of 0.2% formic acidin water-acetonitrile(40∶60,V/V)at a flow rate of 0.4 mL·min-1. The mass spectrometer was operated in a positive electrospray multiple reaction monitoring(MRM) mode. The pharmacokinetic parameters were calculated by Phoenix WinNonlin 6.2. RESULTS The calibration curve was linear over the range10-20 000 μg·L-1 for paclitaxel. Both inter-and intra-day precision(RSD) were less than 8%. The average recovery was between 89.5% and 97.7%. IS normalized-MF was between 0.8887 and 1.033, RSD was less than 4%. This method was applied successfully to quantify paclitaxel level in plasma of 8 patients with advanced solid cancer who received recommended dose of drugs(300 mg·m-2) by intravenous infusion for 3 h. The main pharmacokinetic parameters were as follows:ρmax was(3 872±1 062)μg·L-1, AUC0-∞was(14 603±3 390) μg·h·L-1, t1/2 is(18.3±6.4) h, CL was(22.0±5.0) L·h-1·m-2.CONCLUSION The established LC-MS/MS detection method is sensitive, accurate, and simple for the determination of paclitaxel in human plasma. This method was successfully applied in pharmacokinetic study.
【Key words】 paclitaxel; chromatography,high pressure liquid; tandem mass spectrometry; pharmacokinetics;
- 【会议录名称】 2015年中国药学大会暨第十五届中国药师周论文集
- 【会议名称】2015年中国药学大会暨第十五届中国药师周
- 【会议时间】2015-11-06
- 【会议地点】中国天津
- 【分类号】O657.63;R730.5
- 【主办单位】中国药学会