新一代时间频率基准:NIM4#激光冷却-铯原子喷泉钟

A Time and Frequency Primary Standard of New Generation: NIM4# Laser Cooling-Cesium Atomic Fountain Clock

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 李天初李明寿林平卫黄秉英钱进王平干云清辛明德陈伟亮石春英赵晓慧刘年丰

【Author】 LI Tianchu, LI Mingshou, LIN Pingwei, HUANG Bingying, QIAN Jin, WANG Ping, GAN Yunqing, XIN Mingde, CHEN Weiliang and SHI Chunying National Institute of Metrology, Beijing (100013)

【机构】 中国计量科学研究院

【摘要】 NIM4#冷原子喷泉钟从2003年8月起已经连续稳定运行5个月,在2003年9-12月完成了系统频率偏差评定。本文简要回顾NIM4#钟的喷泉实验和微波锁定,着重报导对NIM4#钟的技术指标实验,实验证明该装置达到频率稳定性8x10-13/τ1/2;复现性5x10-15和系统频率偏移评定不确定度8.5x10-15.

【Abstract】 The NIM4# atomic fountain clock has been operating stably and continually for 5 months since August 2003 and a first run evaluation for systematic frequency deviations was performed during 9-12,2003. We in this paper review briefly the fountain experiments and microwave locking, and report the experiments certifying a stability of 8x10-13/τ1/2, reproducibility of 5 x10-15, and the evaluation of frequency deviations with an uncertainty of 8.5 x10-15.