节点文献
失配滤波器在高频超声成像编码激励技术中的应用
Application of mismatched filters in coded excitation for high-frequency ultrasound imaging
【Author】 WANG Xiaochun,ZHOU Sheng,YANG Jun,JI Jianjun,WANG Yanqun(Biomedical Engineering Institute,Chinese Academy of Medical Sciences,Tianjin,300192,China)
【机构】 中国医学科学院生物医学工程研究所;
【摘要】 在高频超声编码激励成像中,通过使用加权迭代最小二乘法来设计失配滤波器,用于数字脉压旁瓣抑制技术中,并通过优化权系数,可以使得输出信号的旁瓣电平压得很低。实验结果表明,通过迭代,选择适当的权系数使得失配滤波器的输出具有较低的旁瓣电平。
【Abstract】 An iteratively weighted least square method is used in high-frequency ultrasound coded excitation imaging to design mismatched filters which are applicable to sidelobe suppression of digital pulse compression.We make the output sidelobe level very low by optimizing the weighting coefficients.The experiment results show that a lower sidelobe level of the mismatched filter output can be achieved by iteratively choosing the proper weighting coefficients.
【Key words】 ultrasound coded excitation; mismatched filter; sidelobe level;
- 【会议录名称】 第十四届中国科协年会第17分会场:环境危害与健康防护研讨会论文集
- 【会议名称】第十四届中国科协年会第17分会场:环境危害与健康防护研讨会
- 【会议时间】2012-09-08
- 【会议地点】中国河北石家庄
- 【分类号】TN957.51
- 【主办单位】中国科学技术协会、河北省人民政府