节点文献

太阳光度计水汽通道定标研究

Study on water-vapor channel calibration of sun photometer

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 黄冬李新韦玮陈胜利孙凌巴秀天

【Author】 HUANG Dong;LI Xin;WEI Wei;CHEN Shengli;SUN Ling;BA Xiutian;Anhui Institute of Optics and Fine Mechanics, HFIPS, Chinese Academy of Sciences;University of Science and Technology of China;National Satellite Meteorological Center,China Meteorological Administration;Meteorological Bureau of Dunhuang;

【通讯作者】 李新;

【机构】 中国科学院合肥物质科学研究院安徽光学精密机械研究所中国科学技术大学中国气象局国家卫星气象中心敦煌市气象局

【摘要】 太阳光度计水汽测量在气候变化研究、光学卫星辐射定标和真实性检验等领域都有着重要应用。为保证长期观测过程中柱水汽含量的高精度测量,开展了太阳光度计水汽通道的改进Langley法定标研究。首先基于参数模型计算了水汽透过率参数,并利用太阳光度计数据计算了单通道法和双通道法的定标系数,随后将反演的大气柱水汽含量与探空数据进行了对比。研究结果表明,单通道法和双通道法结果较为一致;晴好稳定天气时,本研究获得的定标系数与标准定标系数的相对偏差优于2.4%;采用本研究获得的定标系数,太阳光度计反演的大气柱水汽含量与探空结果相对偏差小于7%。此外,需要注意的是,虽然改进的Langley法定标水汽通道具有较好的适用性,但保证该通道的定标精度需要满足晴好稳定的大气条件,同时需要降低936 nm通道的气溶胶光学厚度计算和大气模式对水汽透过率参数计算的影响。

【Abstract】 Objective Columnar water vapor(CWV) is an important parameter for characterizing the spatial and temporal distribution of atmospheric moisture, and its measurement plays a significant role in the analysis of the earth-atmosphere radiation balance, climate change studies, and satellite radiometric calibration and validation. Sun photometers have the advantages of measuring CWV over the entire atmospheric column, providing high temporal resolution, and enabling continuous daytime observations. However, the calibration coefficient of the 936 nm water-vapor absorption channel may drift with instrument aging and the changing of environmental conditions. To address the limitations of conventional calibration methods, which are time-consuming and difficult to coordinate with continuous observations, this study investigates the field calibration of the 936 nm water-vapor channel of a sun photometer to improve CWV retrieval accuracy and support related operational applications.Methods Based on the observation data from a CE318 sun photometer deployed at the Dunhuang radiometric calibration site, field calibration of the 936 nm water-vapor channel of the sun photometer was carried out using an improved Langley method. First, the principles of the single-channel(water-vapor channel) method and the dual-channel(water-vapor channel and non-water-vapor channel) method were introduced. Then, water-vapor transmittance parameters were calculated using LBLRTM v12.9 and the HITRAN 2016 spectroscopic database to obtain the corresponding model parameters. Subsequently, the observation data acquired under relatively clear-sky conditions on August 8 and 9, 2018, were selected for fitting to derive the calibration coefficient of the 936 nm channel, and the effects of atmospheric stability, water-vapor transmittance parameters, and estimation errors of aerosol optical depth at 936 nm on the calibration results were further analyzed. Finally, the field calibration results were applied to CWV retrieval and validated against radiosonde observations.Results and Discussion The results showed that under the selected clear-sky conditions, both the single-channel and dualchannel methods achieved good fitting performance, with correlation coefficients exceeding 0.99. The calibration coefficients derived from the two methods were in good agreement, with a relative difference of about 1%. Compared to the standard calibration coefficients, the deviations for the calibration coefficients obtained on August 8, 2018 were relatively small, whereas those on August 9 were noticeably larger, indicating that the accuracy of field calibration is sensitive to atmospheric stability during the observation period. Further analysis revealed that the main factors affecting the calibration results include atmospheric stability, the value of the water-vapor transmittance parameter b, and the accuracy of aerosol optical depth estimation at 936 nm. Among these factors, utilizing more realistic atmospheric profiles and more accurate aerosol parameters can help reduce systematic errors. When the calibration coefficients obtained on August 8, 2018 were applied to CWV retrieval, the retrieved results were highly consistent with radiosonde observations, indicating that the field calibration results are reliable under stable atmospheric conditions.Conclusions It is proved in this work that the field calibration method for the 936 nm water-vapor channel of the sun photometer based on the imporved Langley method is feasible. Both the single-channel and dual-channel methods can provide reliable calibration results and meet the basic accuracy requirements for CWV retrieval and related satellite applications. Howvever, the performance of field calibration is strongly influenced by atmospheric conditions, and clear and stable skies are essential for ensuring calibration accuracy. Future work should establish a more representative lookup table of water-vapor transmittance parameters based on seasonal radiosonde profiles over the Dunhuang region, and further optimize the calibration of non-water-vapor channels and aerosol interpolation methods to improve the stability of field calibration and the accuracy of CWV retrieval.

【基金】 国家重点研发计划项目(2018YFB0504600,2018YFB0504604)
  • 【文献出处】 大气与环境光学学报 ,Journal of Atmospheric and Environmental Optics , 编辑部邮箱 ,2026年02期
  • 【分类号】P407
  • 【下载频次】7
节点文献中: 

本文链接的文献网络图示:

本文的引文网络