基于丝网印刷制备高灵敏度脉诊微裂纹传感器

Preparation of Highly Sensitive Pulse Diagnostic Microcrack Sensors Based on Screen Printing

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 刘佳明强明浩黄尧吴大鸣许红孙靖尧

【Author】 LIU Jiaming;QIANG Minghao;HUANG Yao;WU Daming;XU Hong;SUN Jingyao;School of Mechanical and Electrical Engineering, Beijing University of Chemical Technology;State Key Laboratory of Organic-Inorganic Composites, Beijing University of Chemical Technology;

【通讯作者】 孙靖尧;

【机构】 北京化工大学,塑料机械与塑料工程研究所北京化工大学,有机-无机复合材料全国重点实验室

【摘要】 微裂纹传感器具有较好的柔性、优秀的灵敏度和很低的检测限度,可用于检测脉搏等微小的生理信号。使用导电碳油为印刷油墨。采用丝网印刷的方法在硅胶膜上印刷传感器,并将其与柔性PCB板组合将微裂纹传感器用于脉搏信号的测量。测试结果表明,印刷油墨具有较好的流变性能,适用于丝网印刷。对微裂纹传感器性能进行测试,在0~20%应变范围内,微裂纹传感器具有较好的线性度和灵敏度(GF1=200)。在20%~30%应变范围内,传感器具有较好的线性度和更佳的灵敏度(GF2=624)。具有快速的响应时间(80 ms)和恢复时间(160 ms),可以检测微小的拉伸(最低至0.15 mm),和较好的力学稳定性能,在5 000次压力循环过程中响应信号无明显变化。将传感器应用于脉搏信号检测,可清晰将分辨出脉搏的周期、幅度以及变化规律,可应用于脉搏信号的获取和脉象诊断,对脉诊客观化以及中西医结合方面均具有较大价值。

【Abstract】 The microcrack sensor, which possessed good flexibility, excellent sensitivity, and a very low detection limit, could be used to detect minute physiological signals like pulse. The article utilized conductive carbon oil as printing ink. By applying screen printing, the sensor was printed on a silicone rubber membrane and combined with a flexible PCB board for pulse signal measurement. The test results indicated that, the printing ink exhibits excellent rheological properties, suitable for screen printing. When tested for its performance, the microcrack sensor demonstrates exceptional linearity and sensitivity(GF1=200) within the strain range of 0~20%. In the strain range of 20%~30%, it maintained equally impressive linearity and even higher sensitivity(GF2=624). With a rapid response time of 80 ms and recovery time of 160 ms, it could detect minimal stretching(down to 0.15 mm) and exhibited good mechanical stability, with no significant changed in response signals during 5 000 pressure cycles. When applied to pulse signal detection, the sensor could clearly distinguish the pulse cycle, amplitude, and variation patterns. It could be used for pulse signal acquisition and pulse diagnosis, offering great value for the objectification of pulse diagnosis and the integration of traditional Chinese and Western medicine.

  • 【分类号】TP212;TS871.1
  • 【下载频次】20
节点文献中: 

本文链接的文献网络图示:

本文的引文网络