基于YY/T 1719—2023的PET/MR质量控制

Quality Control for PET/MR Based on YY/T 1719—2023

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 张鹏孟祥溪黄正山张璞孙劼李成伟朱建平周海伟梁延延李囡

【Author】 ZHANG Peng;MENG Xiangxi;HUANG Zhengshan;ZHANG Pu;SUN Jie;LI Chengwei;ZHU Jianping;ZHOU Haiwei;LIANG Yanyan;LI Nan;National Institute of Metrology;Key Laboratory of Carcinogenesis and Translational Research(Ministry of Education/Beijing),Key Laboratory for Research and Evaluation of Radiopharmaceuticals (National Medical Products Administration),Department of Nuclear Medicine,Peking University Cancer Hospital&Institute;Shenzhen Ruida Testing Technology Co.,Ltd.;Shanghai United Imaging Healthcare Co.,Ltd.;

【通讯作者】 李囡;

【机构】 中国计量科学研究院北京大学肿瘤医院暨北京市肿瘤防治研究所核医学科,国家药监局放射性药物研究与评价重点实验室,放射性药物研究与试验评价北京市重点实验室,恶性肿瘤发病机制及转化研究教育部重点实验室深圳市瑞达检测技术有限公司上海联影医疗科技股份有限公司

【摘要】 YY/T 1719—2023《正电子发射断层成像及磁共振成像设备通用技术要求》是国内首个PET/MR质控相关行业标准,基于该标准对PET/MR的核心参数检测技术进行研究。以联影uPMR 790 PET/MR为例,使用ACR MRI模体、NEMA IEC躯干模体等对PET/MR的信噪比、均匀性、空间分辨率、配准精度等参数进行性能测试和数据分析,并对测试方法进行探讨研究。所检测的性能参数均在标准要求及制造商提供的正常范围以内,MR部分的信噪比779.23、均匀性91.04%、层厚相对误差4.02%、最大几何畸变0.40 mm、低对比度可见轮辐34;PET部分断层成像灵敏度17.8 cps/kBq、散射分数44%、峰值NECR132.7 kcps、配准精度最大偏差1.4 mm。YY/T 1719—2023中使用的测试方法,大多为国内外较为通行且比较成熟的技术,具备可执行性,但部分参数所用方法相较于国际标准存在一定的滞后性,且在实际操作中具有一定的复杂度。

【Abstract】 Research on the core parameter detection technology of PET/MR based on YY/T 1719—2023 “General requirements of positron emission tomography and magnetic resonance equipment”, which is the first industry standard related to PET/MR quality control in China. Taking uPMR 790 PET/MR as an example, performance tests and data analysis were conducted on PET/MR parameters such as signal-to-noise ratio, uniformity, spatial resolution, and registration accuracy using models such as the ACR MRI phantom and NEMA IEC body phantom, and the testing methods were explored and studied. The results obtained are all within the standard requirements and the normal range provided by the manufacturer, with a SNR of 779.23, uniformity of 91.04%, relative error of layer thickness of 4.02%, maximum geometric distortion of 0.40 mm, low contrast visible spokes of 34 of the MR system; tomography sensitivity of17.8 cps/kBq, scattering fraction of 44%, peak NECR132.7 kcps, and maximum registration accuracy deviation of1.4 mm of the PET system. The testing methods used in YY/T 1719—2023 are mostly commonly used and mature technologies both domestically and internationally, which are executable. However, some of the methods used for certain parameters have some lag compared to international standards and have a certain level of complexity in practical operation.

【基金】 国家重点研发计划“诊疗装备与生物医用材料重点”专项(2022YFC2409402);中国计量科学研究院基本科研业务费(AKYZZ2439)
  • 【文献出处】 计量科学与技术 ,Metrology Science and Technology , 编辑部邮箱 ,2025年10期
  • 【分类号】TP391.41;R817;R445.2
  • 【下载频次】5
节点文献中: 

本文链接的文献网络图示:

本文的引文网络