3D打印金属椎体替代物力学性能评价

Evaluation of Mechanical Properties of Three-Dimensional-Printed Metal Vertebral Body Substitutes

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 董恩纯康建峰孙畅宁李涤尘罗洋王玲栗向东

【Author】 DONG Enchun;KANG Jianfeng;SUN Changning;LI Dichen;LUO Yang;WANG Ling;LI Xiangdong;State Key Laboratory for Manufacturing System Engineering, School of Mechanical Engineering, Xi’an Jiaotong University;National Medical Products Administration (NMPA) Key Laboratory for Research and Evaluation of Additive Manufacturing Medical Devices;School of Mechatronic Engineering and Automation, Foshan University;Tianjin Medical Devices Quality Supervision and Testing Center;Department of Orthopedics, Xijing Hospital, Air Force Medical University;

【通讯作者】 王玲;栗向东;

【机构】 西安交通大学机械工程学院,机械制造系统工程国家重点实验室国家药品监督管理局医用增材制造器械研究与评价重点实验室佛山科学技术学院机电工程与自动化学院天津市医疗器械质量监督检验中心空军军医大学西京医院骨科

【摘要】 目的 对钛网及3D打印金属椎体替代物力学性能进行研究,为临床中人工椎体选择与结构优化设计提供指导。方法 通过压缩力学测试,对钛网及3D打印多孔型、桁架型与拓扑型椎体替代物的等效力学属性与结构破坏形式进行系统研究。结果 钛网等效弹性模量[(2 908.73±287.39) MPa]仅次于拓扑型椎体替代物,但其结构强度与稳定性较差,等效屈服强度[(46.61±4.85) MPa]高于多孔型椎体替代物,且在压缩中率先屈服;多孔型椎体替代物结构强度[(18.14±0.17)~(25.79±0.40) MPa)]不足,难以满足脊柱重建力学要求;桁架型椎体替代物等效弹性模量[(2 477.86±55.19)~(2 620.08±194.36)MPa]与等效屈服强度[(77.61±0.50)~(88.42±1.07) MPa]良好但稳定性不足,在压缩过程中容易出现失稳现象;拓扑型椎体替代物具有最高的等效弹性模量[(3 746.28±183.80) MPa]与等效屈服强度[(177.43±3.82) MPa],可为人工椎体在体服役安全稳定提供更强保障。结论 拓扑优化方法可实现椎体替代物高强度、高稳定性设计,提供更大的设计空间与安全余量,为人工椎体轻量化与新材料设计提供更多可能。

【Abstract】 Objective To study the mechanical properties of titanium mesh and three-dimensional(3D)-printed metal vertebral body substitutes(VBS) to provide guidance for the selection and structural optimization of artificial vertebral implants in clinical practice. Methods The equivalent elastic modulus, equivalent yield strength, and structural failure mode of titanium mesh and 3D-printed porous, truss, and topologically optimized VBS were systematically investigated using compression tests. Results The elastic modulus of the titanium mesh(2 908.73±287.39 MPa) was only lower than that of the topologically optimized VBS. However, their structural strengths and stabilities were inadequate. The yield strength of the titanium mesh(46.61±4.85 MPa) was only higher than that of the porous VBS and it was the first to yield during compression. The porous VBS was insufficient for use as the vertebral implant owing to its poor mechanical strength(18.14±0.17 MPa-25.79±0.40 MPa). The truss VBS had good elastic modulus(2 477.86±55.19 MPa-2 620.08±194.36 MPa) and strength(77.61±0.50 MPa-88.42±1.07 MPa). However, the structural stability of the truss VBS was insufficient, and instability occurred easily during compression. The topologically optimized VBS had the highest elastic modulus(3 746.28±183.80 MPa) and yield strength(177.43±3.82 MPa) among all the tested VBS types, which could provide improved security and stability for artificial vertebral implant in vivo services. Conclusions Topology optimization results in a high strength and high stability VBS design. Moreover, it provides a large design space and great safety margin to provide increased possibilities for lightweight and new material design of future artificial vertebral implants.

【基金】 国家重点研发计划项目(2022YFB4600500);陕西省自然科学基础研究计划项目(2022JQ-378);中国博士后科学基金项目(2020M683458);国家自然科学基金项目(51835010,12202347);广东省重点领域研发计划项目(2018B090906001);中央高校基本科研业务费;陕西省创新团队(2023-CX-TD-17);陕西省科技统筹创新工程计划项目(2016KTCQ01-44)
  • 【文献出处】 医用生物力学 ,Journal of Medical Biomechanics , 编辑部邮箱 ,2024年01期