数值孔径对飞秒激光角膜切削的影响
Effect of Numerical Aperture on Femtosecond Laser Corneal Ablation
【摘要】 飞秒激光得益于其精准、微创等优势被广泛应用于屈光疾病的治疗。在飞秒激光屈光手术中,光学系统的数值孔径是影响手术效果的重要参数。本研究旨在探讨数值孔径对飞秒激光角膜基质切削质量的影响规律,以帮助临床医生更好地选择合适的手术参数。选用0.16、0.30、0.80三种数值孔径进行离体动物角膜的飞秒激光切削实验,并通过气泡尺寸与凋亡细胞比例评估激光切削质量与基质细胞损伤程度。实验结果显示:气泡体积随着数值孔径的增大而减小,高数值孔径下切割更易实现基质层的分离;上述三种数值孔径下的基质细胞损伤比例分别为9.4%、4.9%和1.0%,基质细胞的损伤比例随着数值孔径的增大而明显下降。因此,增大数值孔径有助于提高飞秒激光角膜基质切削的安全性。
【Abstract】 Objective Corneal laser refractive surgery is a method for correcting vision using lasers to reshape the cornea and change its curvature and thickness. Femtosecond laser corneal cutting is widely used in ophthalmic refractive surgery as a precise, minimally invasive, and controllable surgical technique. In femtosecond laser refractive surgery, the numerical aperture of the optical system determines the focal spot size and required single-pulse energy, which are critical parameters that influence the corneal cutting quality.In this study, we built a femtosecond laser surgery system with an adjustable numerical aperture. We investigated the effect of numerical aperture on cutting quality in the corneal stroma by analyzing the differences in bubble morphology, smoothness of flap separation, and proportion of damaged stromal cells. This study aimed to assist clinicians in selecting the appropriate surgical parameters more effectively.Methods Freshly enucleated pig eyeballs and New Zealand white rabbits were selected as experimental subjects. By adjusting the diameter of the incident beam, corneal flaps were formed on the pig eyeballs using a femtosecond laser with numerical aperture values of 0.16, 0.30, and 0.80. The morphology of the bubbles after cutting was recorded, and the smoothness of the separation was observed when the corneal flaps were lifted. Cell damage experiments were conducted by cutting New Zealand white rabbit eyeballs with a femtosecond laser at numerical aperture values of 0.16, 0.30, and 0.80. After creating the flap with the femtosecond laser, the rabbit eyeballs were placed in corneal active medium(DX solution) and incubated at 4 ℃ for 6 h to induce apoptosis in the damaged corneal stromal cells. Subsequently, the rabbit eyeballs were removed and prefixed in a 4% paraformaldehyde(PFA) solution for 2 h.After dewaxing and rehydration, the corneal sections were double-stained with DAPI(4,6-diamidino-2-phenylindole, D9542, SigmaAldrich) and TUNEL(TdT-mediated dUTP nick end labeling, C1088, Beyotime). Finally, the apoptotic cell counts were determined by imaging the sections under a fluorescence microscope.Results and Discussions Under the three different numerical aperture values(0.16, 0.30, and 0.80), as the numerical aperture increases, the volume of the bubbles decreases gradually, and the density of the bubble layer increases(Figs. 5 and 6). This is mainly attributed to the decreasing volume of the focal spot with an increasing numerical aperture, which decreases cavitation bubbles.Corneal flaps formed at a higher numerical aperture are easier to separate. This is primarily because smaller cavitation bubbles result in a denser bubble layers, which facilitates the separation of the interlamellar space with less adhesions between the tissue layers. In the cell damage experiment, as numerical aperture increases, the number of apoptotic cells decreases significantly, as shown in Fig.7. This is attributed to the decreased single-pulse energy and decreased focal spot size associated with an increase in numerical aperture, which results in smaller photodisruption zones and reduced damage to the surrounding tissues. Therefore, increasing the numerical aperture is beneficial for reducing the extent of stromal cell damage.Conclusions The effects of femtosecond laser corneal cutting for different numerical aperture values were investigated experimentally. The morphological differences in cavitation bubbles induced by a femtosecond laser at different numerical aperture values were analyzed, and the ease of separation of the lamellar layers and the extent of cell damage were compared. The results of the experiment show that during femtosecond laser corneal cutting, a higher numerical aperture yields smaller bubbles, denser bubble layers, easier separation of corneal flaps, and lower levels of damage to stromal cells. Therefore, a higher numerical aperture is beneficial in femtosecond laser refractive surgery. Overall, this study provides valuable insights into the effects of numerical aperture on femtosecond laser corneal cutting and highlights the importance of optimizing the numerical aperture to achieve improved treatment outcomes in corneal procedures.
【Key words】 laser technique; femtosecond laser; cornea; numerical aperture; bubble; cell damage;
- 【文献出处】 中国激光 ,Chinese Journal of Lasers , 编辑部邮箱 ,2024年09期
- 【分类号】R779.63;TN249
- 【下载频次】15