高剂量率近距离192Ir治疗源的井型电离室仪器的校准
Calibration of Well Type Ionization Chamber Instruments with High-dose-rate After-loading 192Ir Source
【摘要】 针对后装治疗用192Ir放射源的参考空气比释动能率的量值急需进行溯源。参考国际通例,由指型电离室(PTW-30013)在60Coγ射线和250 k V X射线下的空气比释动能校准因子推导得出192Irγ射线的空气比释动能校准因子,从而测定192Ir放射源参考空气比释动能率标准值,进而完成对井型电离室的参考空气比释动能率校准。通过不确定度评定得出:192Ir放射源的参考空气比释动能率的不确定度为3.6%,井型电离室校准因子的不确定度为3.8%。
【Abstract】 There is an urgent need for traceability of the reference air kerma rate for the192Ir source for high-dose-rate after-loading. Referring to the international general practice,the calibration factor of the192Ir gamma ray is derived from the calibration factor of air kerma of finger ionization chamber under60Co gamma ray and 250 k V X-ray,so as to determine the standard value of the192Ir gamma ray reference air kerma rate,and then the calibration of reference air kerma rate of well type ionization chamber is completed. The results show that the uncertainty of the192Ir gamma ray reference air kerma rate is 3. 6%,and that of calibration factor of well type ionization chamber is 3. 8%.
【Key words】 metrology; reference air kerma rate; brachytherapy; 192Ir source; well type ionization chamber;