节点文献

高纯锗能谱仪系统研制

Development of HPGe Spectrometer System

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 何高魁孙慧斌郝晓勇刘义保代传波邓长明王柱廖辉汪天照胡世鹏邵俊琪张怀强刘海峰杨松贾伟强厉文聪刘洋阙子昂田华阳张向阳

【Author】 HE Gao-kui;SUN Hui-bin;HAO Xiao-yong;LIU Yi-bao;DAI Chuan-bo;DENG Chang-ming;WANG Zhu;LIAO Hui;WANG Tian-zhao;HU Shi-peng;SHAO Jun-qi;ZHANG Huai-qiang;LIU Hai-feng;YANG Song;JIA Wei-qiang;LI Wen-cong;LIU Yang;QUE Zi-ang;TIAN Hua-yang;ZHANG Xiang-yang;China Institute of Atomic Energy;Shenzhen University;East China University of Technology;Wuhan Second Ship Design and Research Institute;China Institute of Radiation Protection;Wuhan University;Hubei Fangyuan Environmental Protection Technology Co Ltd;

【机构】 中国原子能科学研究院深圳大学东华理工大学武汉船舶第二研究所中国辐射防护研究院武汉大学湖北方圆环保科技有限公司

【摘要】 介绍了高纯锗能谱仪系统的研制。项目突破了探测器级高纯锗单晶提纯、高纯锗探测器表面钝化保护、高计数率数字化多道分析器以及无源效率刻度等关键技术,研制出高纯锗能谱仪系统样机。开展了高纯锗能谱仪在核燃料包壳破损在线监测、中高放射性废物桶核素源项无损检测和材料点缺陷特性分析示范应用研究。性能测试表明:85%探测效率的能谱仪的能量分辨率为1.98 keV(1.33 MeV伽马射线),峰康比好于61。

【Abstract】 The development of High Purity Germanium spectrometer system is introduced.This project has broken through the key technologies of detector grade high-purity germanium single crystal purification, high-purity germanium detector surface passivation, high counting-rate digital multi-channel analyzer and source-less efficiency calibration, and successfully developed a highpurity germanium spectrometer system.The demonstration application research of HPGe spectrometer in on-line monitoring of nuclear fuel cladding damage, non-destructive detection of nuclide source term of medium and high-level radioactive waste barrels and analysis of material point defect characteristics were carried out.The performance test shows that the energy resolution of the spectrometer with 85% detection efficiency is 1.98 keV(1.33 MeV gamma ray), and the peak-to-Compton ratio is better than 61.

【基金】 国家重点研发计划重点专项(2017YFF0106500)资助
  • 【文献出处】 核电子学与探测技术 ,Nuclear Electronics & Detection Technology , 编辑部邮箱 ,2021年01期
  • 【分类号】TL814
  • 【被引频次】1
  • 【下载频次】340
节点文献中: 

本文链接的文献网络图示:

本文的引文网络