用于凝血酶原时间检测的多壁碳纳米管增强型电化学传感器

Multi-walled carbon nanotubes enhanced electrochemical sensor for prothrombin time detection

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 李超郭振张威姚佳孔慧严茹红周连群

【Author】 LI Chao;GUO Zhen;ZHANG Wei;YAO Jia;KONG Hui;YAN Ru-hong;ZHOU Lian-qun;University of Science and Technology of China;Key Laboratory of Biomedical Detection Technology,Suzhou Institute of Biomedical Engineering and Technology,Chinese Academy of Sciences;School of Electronic and Information Engineering,Soochow University;Suzhou hospital Affiliated to Nanjing Medical University;

【通讯作者】 周连群;

【机构】 中国科学技术大学中国科学院苏州生物医学工程技术研究所中国科学院生物医学检验技术重点实验室苏州大学电子信息学院苏州科技城医院检验科

【摘要】 针对目前凝血检测仪器存在的成本昂贵操作复杂等问题,制作了一种基于丝网印刷技术的多壁碳纳米管(MWCNTs)增强型电化学传感器用于凝血酶原时间(PT)测量。利用凝血酶切割凝血酶底物实验验证了计时电流法检测凝血酶原时间原理的可行性。接着对血浆凝血酶原时间参数进行测量,并用SYSMEX CS 5100光学凝血仪验证测量结果。经过测试,凝血酶验证性实验中MWCNT型电化学传感器的电流响应强度较普通电化学传感器增加了(36±1)%,重复实验出峰时间和峰值电流变异系数分别为2.99%和3.27%。测试不同血样PT值,能够清晰的显示出区分度,且挑选三组血样进行PT重复实验,出峰时间变异系数分别为2.26%、3.22%和2.96%,实验结果与医院的临床结果线性拟合决定系数R~2为0.986。MWCNTs增强型型电化学传感器用于PT测试具有良好的重复性和一致性,易于批量生产,大大降低了凝血测量的成本,且适合多种场合的测量,在即时检测领域具有极大潜力。

【Abstract】 The determination of coagulation parameters is of crucial significance in the guided treatment of anticoagulant drug users and patients with liver diseases.Considering the challenges associated with commercial coagulation detection instruments such as high cost and complex operation,a MWCNTenhanced screen-printed electrochemical sensor was developed for coagulation measurement.Initially,the measurement of thrombin substrate cleaved by thrombin was used to evaluate the feasibility of prothrombin time(PT)detection by chronoamperometry.Then the plasma PT parameters were measured and the results were validated using the SYSMEX CS 5100 optical coagulator.The response intensity of a MWCNT electrochemical sensor in thrombin validation experiments increase by(36±1)% compared with a general electrochemical sensor and the peak time coefficient of variation and peak current coefficients of variation are 2.99% and 3.27% respectively.The testing of PT values of different blood samples clearly shows discrimination.Three groups of blood samples are selected for repeated PT parameter measurements and the coefficients of variation are 2.26%,3.22%,and 2.96%respectively.The linear fitting decision coefficient R~2 is 0.986 for clinical results.The MWCNT screen-printed electrochemical sensor for PT testing has good repeatability and consistency,is easy to mass produce,reduces the cost of coagulation measurement,and is suitable for measurement in many circumstances.As such,this sensor has great potential in the field of point-of-care testing.

【基金】 国家自然科学基金资助项目(No.51675517,No.61874133);江苏省自然科学基金资助项目(No.BK20160057,No.BK20170428);中国科学院STS项目资助(No.KFJ-STS-SCYD-217);江苏省重点研究开发项目资助(No.BE2018080);中国博士后科学基金资助项目(No.2018T110552,No.2016M601890);江苏省博士后科研计划资助项目(No.1701010A,No.2018K071C);苏州市医疗工程计划资助项目(No.Y851431103、No.Y851391105、No.Y831191103);苏州市科技计划资助项目(No.SS201749);苏州新区医疗卫生科技计划资助项目(No.2016Q006)
  • 【文献出处】 光学精密工程 ,Optics and Precision Engineering , 编辑部邮箱 ,2019年06期
  • 【分类号】TP212.2;TH77
  • 【被引频次】6
  • 【下载频次】202
节点文献中: 

本文链接的文献网络图示:

本文的引文网络