节点文献
气相色谱法测定一次性使用空心纤维血液透析器中N-甲基吡咯烷酮在模拟临床过程中溶出量
The Determination of N-methyl Pyrrolidone in A Single-use Hollow Fiber Blood Dialyzer was Determined by Gas Chromatography
【摘要】 建立血液透析器中N-甲基吡咯烷酮(NMP)在模拟临床试验过程中溶出量的测定方法,选择乙醇水混合溶液作为浸提模拟溶液,采用气相色谱(GC)检测。结果表明:该方法 NMP的浓度在10~100μg/L范围内线性关系良好(r=0.9993),平均回收率为109.31%(RSD=1.74%,n=9)。方法简单快速,重现性好,回收率高,精密度好,干扰少,定性定量准确,可用于一次性使用空心纤维血液透析器中NMP溶出量的日常监管检测。
【Abstract】 Gas chromatography(GC) is developed in the determination of blood dialyzer N-methyl pyrrolidone(NMP) in simulated the dissolubility of the process of clinical trials, choose ethanol water mixture as leaching solution. The results showed that the concentration of NMP was good(r=0.9993) and the average recovery rate was 109.31%(RSD=1.74%, n=9). This method is simple and fast, good reproducibility, high recovery rate, good precision, less disturbance, qualitative quantitative, accurate and can be used for a single-use hollow fiber blood dialyze dissolubility of NMP daily supervision detection.
- 【文献出处】 中国医疗器械信息 ,China Medical Device Information , 编辑部邮箱 ,2018年07期
- 【分类号】R197.39
- 【被引频次】2
- 【下载频次】127