节点文献
眼晶体热释光剂量H_p(3)测量系统校准方法
The calibration method of the H_p(3) measurement system for the thermoluminescence of the eye lens
【摘要】 通过文献调研,依托防护水平X、γ辐射空气比释动能率标准装置,参照JJG 593-2016《个人和环境监测用X、γ辐射热释光剂量测量系统》,采用高20 cm、直径20 cm的正圆柱体型的有机玻璃模体和德国联邦物理技术研究院(PTB)研究发布的空气比释动能K_a与眼晶体剂量当量H_p(3)的转换系数,实验验证了眼晶体热释光剂量计的校准方法。
【Abstract】 Through literature research, relying on the protection level of X and gamma radiation air kerma rate standard device, according to JJG 593-2016 Thermoluminescence Dosimetry Systems Used in Personal and Environmental Monitoring for X and γ Radiation, A cylindrical with high 20 cm, the diameter of 20 cm organic glass mould body is used with the conversion coefficient of the air kerma K_a to lens dose equivalent H_p(3) studied by the Physikalisch-technische-Bundesanstalt(PTB), The calibration method of the thermoluminescence dosimeter for the eye lens was verified by the experiment.
- 【文献出处】 上海计量测试 ,Shanghai Measurement and Testing , 编辑部邮箱 ,2018年01期
- 【分类号】R144
- 【下载频次】80