An absolute calibration setup for laser-induced breakdown spectroscopy detection system is built based on the radiometry theory.A tungsten halogen lamp equipped with an ultraviolet glass filter and a fused silica diffuser is used as standard light source for the calibration.Absolute calibration of a laser-induced breakdown spectroscopy detection system with a Czerny-Turner spectrometer in the ultraviolet range is performed.The absolute spectral response of the system in the range of 310~385 nm is obtained w...
【基金】
National Key Scientific Instrument and Equipment Development Projects of China(No.2014YQ120351);
National Natural Science Foundation of China(No.11704372);
Anhui Provincial Natural Science Foundation(No.1708085QF130)
【更新日期】
2018-10-09
【分类号】
O657.31;TN249
【正文快照】
Foundation item:National Key Scientific Instrument and Equipment Development Projects of China(No.2014YQ120351),NationalNatural Science Foundation of China(No.11704372)and Anhui Provincial Natural Science Foundation(No.1708085QF130)0 IntroductionLaser-ind