评价血细胞比容、维生素C和半乳糖等干扰因素对3种便携式血糖仪的影响

Evaluation on influence of interference factors HCT,vitamin C and galactose on three different portable blood glucose meters

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 孙德华毛欣茹郑磊艾育华李文源

【Author】 SUN Dehua;MAO Xinru;ZHENG Lei;AI Yuhua;LI Wenyuan;Department of Clinical Laboratory,Nanfang Hospital,Southern Medical University;Department of Information,Nanfang Hospital,Southern Medical University;General Administration Office,Nanfang Hospital,Southern Medical University;

【机构】 南方医科大学南方医院检验科南方医科大学南方医院信息科南方医科大学南方医院院办公室

【摘要】 目的评价血细胞比容(HCT)、维生素C(Vc)和半乳糖等干扰因素对3种便携式血糖仪的影响,为临床选择合适的血糖仪提供一定参考。方法选取该院检验科肝素锂抗凝的静脉血标本20份用于血糖仪准确度验证,选取2名健康志愿者的空腹静脉全血标本各5份用于干扰物试验,参考罗氏C501全自动生化分析仪的检测结果,对Nova StatStrip Xpress血糖仪、拜安进血糖仪、罗氏卓越血糖仪进行准确度验证,评价HCT、Vc和半乳糖对以上3种便携式血糖仪检测结果的影响。结果 Nova StatStrip Xpress血糖仪和罗氏卓越血糖仪检测准确度均符合ISO15197∶2013标准要求,拜安进血糖仪检测准确度仅符合ISO15197∶2003标准要求。在不同水平HCT、Vc、半乳糖干扰下,Nova StatStrip Xpress血糖仪检测结果均符合ISO15197∶2013标准要求;拜安进血糖仪检测低浓度血糖时受到10mg/dL Vc干扰,其余检测结果均符合ISO15197∶2003标准要求;罗氏卓越血糖仪对HCT的抗干扰性能符合ISO15197∶2013标准要求,但受Vc和半乳糖干扰较大。结论 3种便携式血糖仪的准确度和抗干扰性能不同,临床应根据患者实际情况选择合适的血糖仪,确保血糖检测的准确性。

【Abstract】 Objective To evaluate the influence of hematocrit(HCT),vitamin C(Vc)and galactose on three portable blood glucose meters to provide some reference for clinical selection of appropriate blood glucose meter.Methods 20 heparin anticoagulant venous blood samples in the clinical laboratory department of Nanfang Hospital were selected for verifying the accuracy of blood glucose meter.2 healthy volunteers were selected for collecting 5 whole blood samples in interferent test.Referring to the detection results of Roche automatic biochemical analyzer,the Nova StatStrip Xpress glucometer,Bayer Contour Plus glucometer and Roche Accu-chek Performa blood glucose meter were performed the accuracy verification.The influence of HCT,Vc and galactose on the detection results of above three portable blood glucose meters were evaluated.Results The detection accuracy of Nova StatStrip Xpress and the Roche Accu-Chek glucometers all conformed to the requirements of ISO15197∶2013 standards.The detection accuracy of Bayer Contour Plus glucometer only conformed to the requirements of ISO15197∶2003 standards.Under the interference of different levels of HCT,Vc and galactose,the detection results of Nova StatStrip Xpress glucometer conformed to the requirements of ISO15197:2013 standards;when detecting low concentration of blood glucose,the Bayer Contour Plus glucometer was interfered by 10 mg/dL Vc,and other detection results conformed to the requirements of ISO15197∶2003 standards;the anti-interference performance of Roche Accu-chek Performa glucometer conformed to the requirements of ISO15197∶2013 standards,but the Vc and galactose interference was relatively great.Conclusion The accuracy and anti-interference performance have difference among the three portable blood glucose meters.Clinic should select suitable blood glucose meter according to the actual situation of the patients to ensure the accuracy of blood glucose detection.

【基金】 2016年广东省级财政资金互联网+领域项目[粤财工(2016)187号];2013年度广东省信息产业发展专项物联网应用扶持项目[粤经信信息函(2014)975号]
  • 【文献出处】 国际检验医学杂志 ,International Journal of Laboratory Medicine , 编辑部邮箱 ,2018年06期
  • 【分类号】R446.11
  • 【被引频次】4
  • 【下载频次】227
节点文献中: 

本文链接的文献网络图示:

本文的引文网络