节点文献

低温浓缩-气质联用分析污染源废气中环氧氯丙烷

Determination of Epichlorohydrin Emissions from Stationary Sources Using Cryogenic Preconcentration Followed by GC/MS

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 宋晓娟李海燕尹明明马玉琴

【Author】 SONG Xiao-juan;LI Hai-yan;YIN Ming-ming;MA Yu-qin;Lianyungang Environmental Monitoring Center;

【机构】 连云港市环境监测中心站

【摘要】 建立了气袋采样-液氮低温浓缩-气质联用分析污染源废气中环氧氯丙烷的方法,优化了相关监测参数。优化后的方法在环氧氯丙烷质量浓度10~500μg/m~3范围内线性良好,相关系数>0.999,加标回收率为72.1%~85.7%,相对标准偏差<5%,气体进样量为5.0 mL时,检出限为4.37μg/m~3,可以满足废气中环氧氯丙烷监测的要求。

【Abstract】 A method based on cryogenic pre-concentration followed by GC/MS was established for the determination of epichlorohydrin emissions from stationary sources. A cryogenic pre-concentrator,which acts as autosampler,was used to concentrate epichlorohydrin collected in gas bags. Relevant parameters for cryogenic pre-concentration were optimized. After pre-concentrated,epichlorohydrin was introduced into GC/MS to undergo qualitative and quantitative analysis,and appraisal of the whole method was then carried out from different respects. Good linearity was obtained within 10 ~ 500 μg/m~3,and the correlation coefficient was greater than 0. 999. The average recoveries ranged from 72. 1% ~ 85. 7%. Meanwhile,the method was stable enough to realize extensive use with RSD lower than 5%. The detection limit was 4. 37 μg/m~3 based on an injection volume of 5 mL,which was sensitive enough to realize waste gas monitoring.

【关键词】 低温浓缩气质联用环氧氯丙烷
【Key words】 Cryogenic pre-concentrationGC/MSEpichorohydrin
  • 【文献出处】 环境监控与预警 ,Environmental Monitoring and Forewarning , 编辑部邮箱 ,2017年02期
  • 【分类号】X831;O657.63
  • 【被引频次】2
  • 【下载频次】57
节点文献中: 

本文链接的文献网络图示:

本文的引文网络