氡钍分析仪水氡观测最佳工作状态

The application of the new radon-thoron analysis instrument in dissolved radon monitoring

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 樊春燕陈华静王艳

【Author】 Fan Chunyan;Chen Huajing;Wang Yan;China Earthquake Networks Center;Hebei Earthquake Agency;

【机构】 中国地震台网中心河北省地震局

【摘要】 中国水氡观测历经50余年,观测仪器老化严重。2011年至今分别引入新生产的3批次64套FD-125型氡钍分析仪。经过仪器测试、地震台站运行以及多次现场实验,发现:该仪器定标器阈值不应按照地震水文地球化学观测技术规范选定的2 V,需选择0.7 V甚至更低阈值,才能保证测值的可靠稳定,保障水氡观测继续发挥防灾减灾作用。

【Abstract】 Radon is inert gas of radioactive decay arising from the rock and soil of crust, which mainly contain the precursors of earthquakes. The problems of instruments aging in dissolved radon monitoring have become more serious in the near days during the continued using over 40 years. 64 sets of FD-125 radon and thorium analytical instrument have been introduced by the projects of background field and Instrument renewal and transformation. We carried on several factory testing and filed experiments in observation stations. The test results show that, on the condition of a relatively low threshold(eg, 0.5 V, or even lower) of the scalar, the results are more stable and reliable. In the conditional case, after choosing suitable threshold, the measurement of high-voltage plateau curve may carry out using radium source or radon gas source. In this process, special attention is being given to avoid light in daily maintenance.

【基金】 全国地下流体物理量观测台网产出(项目编号:1640410320103)
  • 【文献出处】 地震地磁观测与研究 ,Seismological and Geomagnetic Observation and Research , 编辑部邮箱 ,2017年03期
  • 【分类号】P315.62
  • 【被引频次】2
  • 【下载频次】90
节点文献中: 

本文链接的文献网络图示:

本文的引文网络