节点文献

顶空进样GC-SIM-MS测定硝唑尼特中溶剂残留甲醇含量

Determination of Methanol Solvent Residue in Nitazoxanide with Headspace GC-SIM-MS

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 吴清盛茅小燕孙丽东

【Author】 WU Qing-sheng;MAO Xiao-yan;SUN Li-dong;Zhejiang Yuying Vocational and Technology College;

【机构】 浙江育英职业技术学院

【摘要】 利用气相色谱-选择离子监测质谱联用仪(GC-SIM-MS)测定硝唑尼特样品中溶剂残留甲醇的含量,顶空进样,温度为70℃,时间为20 min.SIM定量离子m/z 32,参考离子m/z 31,30和29.甲醇质量浓度在0.5~100μg/m L内呈良好的线性(R~2=0.999 92).方法相对标准偏差为2.49%~4.85%,样品加标回收在94.5%~102.8%之间,检测限为0.5μg/m L,符合定量分析的要求.

【Abstract】 The determination of methanol solvent residues in the samples of nitazoxanide using gas chromatography-selected ion monitoring-mass spectrometry( GC-SIM-MS) has been curried out. A peak area standard curve method is used as the quantitative method. The heating temperature of the sample was 70 ℃,with an equilibrium time of 20 min,SIM m / z 32,m / z 31,30,and 29. The concentration of toluene was linear in the range of 0. 5 ~ 100 μg / mL( R~2= 0. 999 92). The relative standard deviation of the method was 2. 49% ~ 4. 85%,the recovery was between 94. 5% ~ 102. 8%,the detection limit was 0. 5 μg / m L,which were in accordance with the requirement of quantitative analysis.

  • 【文献出处】 分析测试技术与仪器 ,Analysis and Testing Technology and Instruments , 编辑部邮箱 ,2016年02期
  • 【分类号】O657.63;TQ465
  • 【被引频次】2
  • 【下载频次】99
节点文献中: 

本文链接的文献网络图示:

本文的引文网络