节点文献
Application of a BC501A Liquid Scintillation Detector with a Gain Stabilization System on the EAST Tokamak
【摘要】 A 2"×2"BC501A liquid scintillation detector with a gain stabilization system is developed and applied to neutron andγ-ray measurement on the EAST tokamak.Energy calibration of a liquid scintillator using a fast coincidence method is presented and compared with the Monte Carlo simulation.Determination of the proton light output function of the BC501A is presented.Results from dedicated experiments with an Am-Be neutron source,γsource and quasi-monoenergetic neutron beams,and from measurements on EAST tokamak are presented and discussed.
【Abstract】 A 2"×2" BC501 A liquid scintillation detector with a gain stabilization system is developed and applied to neutron and γ-ray measurement on the EAST tokamak. Energy calibration of a liquid scintillator using a fast coincidence method is presented and compared with the Monte Carlo simulation. Determination of the proton light output function of the BC501 A is presented. Results from dedicated experiments with an Am-Be neutron source, γ source and quasi-monoenergetic neutron beams, and from measurements on EAST tokamak are presented and discussed.
【Key words】 gain stabilization; BC501A; neutron spectrometer; pulsed light source; EAST;
- 【文献出处】 Plasma Science and Technology ,等离子体科学和技术(英文版) , 编辑部邮箱 ,2016年01期
- 【分类号】TL812
- 【被引频次】2
- 【下载频次】22