【摘要】 目的对本院2014年1月-12月血培养阳性分离菌及假阳性的报警时间进行统计分析。方法采用Bac T/ALERT 3D 120进行血培养,记录并传输阳性瓶报警时间,VITEK-2 Compact对分离菌进行鉴定和药敏试验。采用Excel统计处理数据。结果本院血培养分离的分离菌总体平均阳性报警时间为27.6 h,中位数为19.6 h,94%以上的阳性血培养瓶在72 h内报警;其中平均阳性报警时间和中位数时间最短的为链球菌(12.0 h/11.4 h),最长的为阳性杆菌(42.7 h/40.5 h);阳性报警时间最短(2 h)的为大肠埃希菌、肺炎克雷伯菌、无乳链球菌和黏质沙雷菌,阳性报警时间>120 h的为表皮葡萄球菌(125.5 h)、近平滑念珠菌(130.5 h和131.7 h)和少动鞘氨醇单胞菌(137.1 h)。假阳性血培养的平均阳性报警时间为69.4 h,中位数时间为61.0 h。结论通过对血培养报警时间的统计分析,可对阳性分离菌的种类进行初步预测,并可辅助假阳性的判断。
【Abstract】 Objective To analyze the positive alarm time of positive pathogen isolates and false positive in blood culture during January- December in 2014. Methods Blood specimens were cultured by Bac T / ALERT 3D 120,and the positive alarm time were recorded and uploaded. Pathogens isolated from blood cultures were identified by VITEK- 2 Compact. The data were analyzed by Excel. Results In 2014,the average positive alarm time of all pathogens isolated from blood cultures was 27. 6 h,and the median alarm time was 19. 6 h. 94% of the isolates alarmed within 72 h. The average and median positive alarm time of Streptococcus spp( 12. 0 h /11. 4 h) was the shortest of all species,and Gram positive bacilli( 42. 7 h /40. 5 h) was the longest. The fastest positive alarm time was 2 h,including Escherichia coli,Klebsiella pneumonia,Streptococcus agalactiae and Serratia marcescens. The isolates alarming time more than 120 h included Staphylococcus epidermidis( 125. 5 h),Candida parapsilosis( 130. 5 h and 131. 7 h) and Sphingomonas paucimobilis( 137. 1 h). The average positive alarm time of false positives in blood cultures was 69. 4 h,and the median alarm time was 61. 0 h. Conclusion The positive alarm time can be used to forecast the species of pathogens isolated from blood cultures,and it is useful to determine the false positive in blood culture,too.
【Key words】 Blood culture; Positive alarm time; Pathogen; False positive;
- 【文献出处】 中国卫生检验杂志 ,Chinese Journal of Health Laboratory Technology , 编辑部邮箱 ,2015年21期
- 【分类号】R446.5
- 【被引频次】1
- 【下载频次】185