节点文献
视频胶囊内窥镜的改进设计与实验
Improvement and Experiment of the Video Capsule Endoscope
【摘要】 为延长视频胶囊内窥镜工作时间,提高体外接收视频质量,需降低胶囊内射频电路模块的功耗以改进发射电路的发射效果.针对上述问题,本文设计了胶囊内的射频电路模块和螺旋形发射天线,并使用ADS2011及HFSS对其进行仿真,最后通过实验对该设计进行验证.内窥镜胶囊进入活体动物(猪)胃肠道后,体外便携接收器接收到的视频图像质量较高,证明该改进设计可行.结果表明:当射频电路板的直径为9.8 mm,射频模块的工作频率为1.0GHz,功耗为15.9 m W时,胶囊的工作时间为9.0 h,视频数据的无线传输速率达到30 frame/s(320×240 pixel/frame).
【Abstract】 To prolong the w orktime of the video capsule endoscopy and improve the quality of the video,low ering the pow er of the embedded radio circuit module inside the capsule is needed,w hich helps improve the radio effect. The embedded radio circuit module and helix antenna inside the endoscopy are designed,ADS2011 and HFSS are applied in simulation,and finally the experiments validate the design. The high quality video after the video capsule endoscopy entered the gastrointestinal( GI) tract of the live animal( pig) validated the feasibility of the design. The result indicates that the w orktime of the endoscope is 9. 0 h and the w ireless transmitting rate reaches 30 frame / s( 320 × 240 pixel / frame) w hen the diameter of transmission circuit board is 9. 8 mm,w ork frequency is 1. 0 GHz,and pow er is 15. 9 m W.
【Key words】 capsule endoscopy; video capsule; RF circuit; helix antenna;
- 【文献出处】 电子学报 ,Acta Electronica Sinica , 编辑部邮箱 ,2015年03期
- 【分类号】TH776.1
- 【被引频次】6