Flash序列低电压小儿肺CT扫描的可行性研究

Feasibility study on application of low tube voltage using Flash mode of DSCT in pediatric chest CT scan

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 唐言柳澄孙丛郭传超毕万利

【Author】 TANG Yan;LIU Cheng;SUN Cong;GUO Chuan-chao;BI Wan-li;School of Medicine,Shandong University,Shandong Medical Imaging Research Institute;Taishan Medical University;

【机构】 山东大学医学院山东省医学影像学研究所泰山医学院

【摘要】 目的对三种不同管电压条件下的Flash模式小儿肺CT图像质量进行评价,探讨低电压扫描的可行性。方法前瞻性分析93例患儿胸部双源CT Flash模式检查资料,按照不同管电压随机分为三组(组1:80kV,34例;组2:100kV,29例;组3:120kV,30例)。两位放射科医师独立对图像质量采用4分制评分;记录每组有效mAs、CTDlvol和DLP,测量图像密度及噪声,计算信噪比(SNR)、有效辐射剂量(ED)。结果①三组图像质量得分〔(3.12±0.640)、(3.69±0.471)、(3.80±0.407)〕及SNR〔(4.25±0.945)、(5.46±1.790)、(5.88±1.665)〕在组1、2和组1、3间均有统计学差异(P<0.05),组2、3间无统计学差异(P>0.05);②三组间的CTDlvol〔(1.05±0.056)mGy、(2.14±0.306)mGy、(2.63±0.697)mGy〕及ED〔(0.69±0.086)mSv、(1.36±0.250)mSv、(1.84±0.808)mSv〕均具有统计学差异(P<0.05)。结论 100kV下CT Flash模式较80kV和120kV具有综合优势,在降低辐射剂量的同时能够保证高图像质量,对小儿肺部检查,尤其是小气道、肺间质等细微结构的观察具有显著临床意义。

【Abstract】 Objective To evaluate the impact of three different tube voltages on image quality and radiation dosage in pediatric chest examination with Flash mode(subsecond high-pitch scan mode,HPM=3.0)of dual-source computed tomography(DSCT)and to analyze the feasibility of the low-voltage CT scan.Methods A total of 93patients undergoing chest scans with Flash mode of DSCT were prospectively enrolled in this study and divided into three groups equally according to different tube voltages(group 1:80kV,34cases;group 2:100kV,29cases;group 3:120kV,30cases).The image quality was evaluated by two radiologists using 4-point scores independently.Image attenuation and noise were measured.Signal-to-noise-ratio(SNR)and effective dose were calculated.The effective mAs,CTDlvol and DLP were recorded.Results The average image quality score of the three groups was(3.12±0.640),(3.69±0.471)and(3.80± 0.407).The average SNR was(4.25±0.945),(5.46±1.790)and(5.88±1.665).There were significant differences in score and SNR between group 1and 2,group 1and 3(P<0.05).There was no difference between group 2and 3(P> 0.05).The average CTDlvol was(1.05±0.056)mGv,(2.14±0.306)mGv and(2.63±0.697)mGv.The average ED was(0.69±0.086)mSv,(1.36±0.250)mSv and(1.84±0.808)mSv.There were significant differences in CTDlvol and ED among three groups(P <0.05).Conclusion Flash mode of DSCT under 100kV has comprehensive advantages in reducing radiation dose while ensuring high image quality,and is of great clinical application value in the pediatric chest CT scan,especially in the observation of fine structures,such as small airways,interstitial tissues,and so on.

  • 【文献出处】 医学影像学杂志 ,Journal of Medical Imaging , 编辑部邮箱 ,2014年04期
  • 【分类号】R816.41
  • 【被引频次】7
  • 【下载频次】67
节点文献中: 

本文链接的文献网络图示:

本文的引文网络