节点文献

Performance simulation and structure design of Binode CdZnTe gamma-ray detector

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 牛莉博李玉兰张岚傅楗强江灏何彬李元景

【Author】 NIU Li-Bo;LI Yu-Lan;ZHANG Lan;FU Jian-Qiang;JIANG Hao;HE Bin;LI Yuan-Jing;Department of Engineering Physics,Tsinghua University;Key Laboratory of Particle & Radiation Imaging(Tsinghua University)Ministry of Education;Xi’an Research Institute of Hi-Tech Hongqing Town;Nuclear Company Limited;

【机构】 Department of Engineering Physics,Tsinghua UniversityKey Laboratory of Particle & Radiation Imaging(Tsinghua University)Ministry of EducationXi’an Research Institute of Hi-Tech Hongqing TownNuclear Company Limited

【摘要】 A new electrode structure CdZnTe(Cadmium Zinc Telluride) detector named Binode CdZnTe has been proposed in this paper.Together with the softwares of MAXWELL,GEANT4,and ROOT,the charge collection process and its gamma spectrum of the detector have been simulated and the detector structure has been optimized.In order to improve its performance further,Compton scattering effect correction has been used.The simulation results demonstrate that with refined design and Compton scattering effect correction,Binode CdZnTe detectors is capable of achieving 3.92%FWHM at 122 keV,and 1.27%FWHM at 662 keV.Compared with other single-polarity(electron-only) detector configurations,Binode CdZnTe detector offers a cost effective and simple structure alternative with comparable energy resolution.

【Abstract】 A new electrode structure CdZnTe(Cadmium Zinc Telluride) detector named Binode CdZnTe has been proposed in this paper.Together with the softwares of MAXWELL,GEANT4,and ROOT,the charge collection process and its gamma spectrum of the detector have been simulated and the detector structure has been optimized.In order to improve its performance further,Compton scattering effect correction has been used.The simulation results demonstrate that with refined design and Compton scattering effect correction,Binode CdZnTe detectors is capable of achieving 3.92%FWHM at 122 keV,and 1.27%FWHM at 662 keV.Compared with other single-polarity(electron-only) detector configurations,Binode CdZnTe detector offers a cost effective and simple structure alternative with comparable energy resolution.

  • 【文献出处】 Nuclear Science and Techniques ,核技术(英文版) , 编辑部邮箱 ,2014年01期
  • 【分类号】TL81
  • 【被引频次】6
  • 【下载频次】93
节点文献中: 

本文链接的文献网络图示:

本文的引文网络