节点文献
便携式代谢率检测系统的误差分析及改进
The Error Analysis and Improvement of Portable Metabolic Rate Detection System
【摘要】 通过对便携式代谢率检测系统误差模型进行蒙特卡罗仿真,得到最终误差指标,通过改进测试电路和多传感器数据融合,在非标定条件下实现了仪器的代谢率测试准确度要求。针对温度传感器测试模块进行了误差分析和改进,精度达到0.15℃(RMS)。通过极大似然估计对改进电路进行数据融合,保证整个测试过程中温度测试误差在0.124℃(RMS)以内,满足了批量生产的高准确度要求,避免了标定过程,降低了生产成本。
【Abstract】 In this paper,utilizing Monte Carlo simulation on the error model of portable metabolic rate detection system,obtained the final error indicators. Improved test circuit and used multi-sensor data fusion to achieve metabolic rate test instrument accuracy requirements in non-calibrated conditions. Through sensor error analysis and improvement,the accuracy of temperature test module of the project reached 0. 15 ℃( RMS). In addition,through the maximum likelihood estimate of the improvement circuit for data fusion,it can ensure that the test error is 0. 124 ℃( RMS)or less throughout the temperature testing process,meet the requirements of high accuracy of mass index,avoided the calibration process,reduced the cost of production.
【Key words】 metabolic rate detection; error statistical analysis; data fusion; temperature sensor;
- 【文献出处】 传感技术学报 ,Chinese Journal of Sensors and Actuators , 编辑部邮箱 ,2014年11期
- 【分类号】TP212;TP202
- 【被引频次】2
- 【下载频次】105