Determination of trace elements in high purity nickel by high resolution inductively coupled plasma mass spectrometry

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 聂西度梁逸曾唐有根谢华林

【Author】 NIE Xi-du1,2,LIANG Yi-zeng1,TANG You-gen1,XIE Hua-lin3 1.School of Chemistry and Chemical Engineering,Central South University,Changsha 410083,China;2.Department of Material and Chemical Engineering,Hunan Institute of Technology,Hengyang 421002,China;3.College of Chemistry and Chemical Engineering,Yangtze Normal University,Fuling 408100,China

【机构】 School of Chemistry and Chemical Engineering,Central South UniversityDepartment of Material and Chemical Engineering,Hunan Institute of TechnologyCollege of Chemistry and Chemical Engineering,Yangtze Normal University

【摘要】 The contents of Mg,Al,Si,Ti,Cr,Mn,Fe,Co,Cu,Ga,As,Se,Cd,Sb,Pb and Bi in high purity nickel were determined by high resolution inductively coupled plasma mass spectrometry(HR-ICP-MS).The sample was dissolved in HNO3 and HCl by microwave digestion.Most of the spectral interferences could be avoided by measuring in the high resolution mode.The matrix effects because of the presence of excess HCl and nickel were evaluated.Correction for matrix effects was made using Sc,Rh and Tl as internal standards.The optimum conditions for the determination were tested and discussed.The detection limits range from 0.012 to 1.76 μg/g depending on the type of elements.The applicability of the proposed method is also validated by the analysis of high purity nickel reference material(NIST SRM 671).The relative standard deviation(RSD) is less than 3.3%.Results for determination of trace elements in high purity nickel were presented.

【Abstract】 The contents of Mg,Al,Si,Ti,Cr,Mn,Fe,Co,Cu,Ga,As,Se,Cd,Sb,Pb and Bi in high purity nickel were determined by high resolution inductively coupled plasma mass spectrometry(HR-ICP-MS).The sample was dissolved in HNO3 and HCl by microwave digestion.Most of the spectral interferences could be avoided by measuring in the high resolution mode.The matrix effects because of the presence of excess HCl and nickel were evaluated.Correction for matrix effects was made using Sc,Rh and Tl as internal standards.The optimum conditions for the determination were tested and discussed.The detection limits range from 0.012 to 1.76 μg/g depending on the type of elements.The applicability of the proposed method is also validated by the analysis of high purity nickel reference material(NIST SRM 671).The relative standard deviation(RSD) is less than 3.3%.Results for determination of trace elements in high purity nickel were presented.

【基金】 Project(21075138) supported by the National Natural Science Foundation of China;Project(cstc2011jjA0780) supported by Natural Science Foundation of Chongqing City,China;Project(KJ121311) supported by Educational Commission of Chongqing City of China
  • 【文献出处】 Journal of Central South University ,中南大学学报(英文版) , 编辑部邮箱 ,2012年09期
  • 【分类号】TG115.33
  • 【被引频次】15
  • 【下载频次】84
节点文献中: 

本文链接的文献网络图示:

本文的引文网络