高效液相色谱法测定白消安血药浓度
Determination of busulfan in human plasma by HPLC
【摘要】 目的 建立测定人血浆中白消安浓度的柱前衍生高效液相色谱法。方法 以1,5-戊二醇二甲磺酸酯为内标,血浆样品经二乙基二硫代氨基甲酸钠衍生化处理后,以乙腈-0.7%冰醋酸水溶液(69:31)为流动相,流速1.0 ml·min-1,采用Xterra?RP18色谱柱分离,在280nm波长下进行检测。结果 白消安线性范围为0.20~4.00 mg·L-1,以加权最小二乘法计算得到回归方程为y=2.56x+16.7(r=0.999 3,n=7),提取回收率81.01%~82.72%,方法回收率96.50%~97.90%,日内、日间精密度(RSD)均小于10%。最低血浆检测质量浓度为0.080 mg·L-1。结论 该方法准确、灵敏,适用于白消安血浆浓度测定及其药代动力学研究。
【Abstract】 Objective To establish a method for determination of plasma busulfan concentration by pre-column derivatization pre-high performance liquid chromatography.Methods 1,5-diol dimethyl sulfonate was used as internal standard,plasma samples were processed by derivatization with acetonitrile-0.7% acetic acid(69:31) as mobile phase,at a flow rate of 1 ml·min-1,Xterra? RP18 column(5 μm,4.6 mm ×250 mm) was used,at 280 nm wavelength detection.Results Busulfan plasma concentration of the linear range of 0.20-4.00 mg·L-1,computing the weighted least squares regression equation for the y=2.56x+16.7(r=0.999 3,n=7).The extraction recovery was81.01%-82.72% and the method recovery was 96.50%-97.90%.The intra-and inter-day RSD were less than10%.Minimum plasma concentration of detection was 0.080 mg·L-1.Conclusion This method is accurate,sensitive and suitable for determination of busulfan plasma concentration and pharmacokinetic studies.
【Key words】 high-performance liquid chromatography; busulfan; plasma concentration; of drug;
- 【文献出处】 徐州医学院学报 ,Acta Academiae Medicinae Xuzhou , 编辑部邮箱 ,2012年05期
- 【分类号】R969.1;O657.72