眼反应分析仪在临床眼压测量中的应用
Application of ocular response analyzer in the intraocular pressure measurement
【摘要】 目的比较眼反应分析仪(ORA)与压平式(GAT)、非接触式眼压计(NCT)眼压测量的差异。方法对正常组、青光眼组及屈光术后组(分别为168眼、62眼、102眼)行ORA测量,正常组及屈光术后组行非接触式眼压计测量,青光眼组行压平眼压计测量,各组均用超声角膜测厚仪测量中央角膜厚度(CCT)。结果正常组、青光眼组及屈光术后组平均ORA测量值分别为(18.33±4.00)mmHg、(19.91±4.44)mmHg、(16.69±5.24)mmHg,NCT/GAT值分别为(17.44±3.53)mmHg、(18.90±3.74)mmHg、(13.91±4.99)mmHg),各组平均ORA测量值均高于NCT/GAT值;正常组及屈光术后组NCT值与CCT值存在中度相关性(r分别为0.466、0.472),青光眼组GAT值与CCT值存在中度相关性(r为0.434),ORA测量值在各组中与CCT均仅存在低度相关性(r分别为0.258、0.221、0.249)。结论与NCT及GAT比较,ORA测量值较少受角膜因素影响,能提供更接近真实眼压的测量结果。
【Abstract】 Objective To compare ocular response analyzer(ORA),goldmann applanation tonometer(GAT)and non-contact tonometer(NCT)in the intraocular pressure measurement.Methods The patients were recruited as the normal group(168 eyes),the glaucoma group(62 eyes) and the post-refractive-surgery group(102 eyes).The intraocular pressure and the central corneal thickness(CCT) of all patients were measured by ORA and A-type ultra-sonography.The intraocular pressure of the normal group and the post-refractive-surgery group was measured by NCT and GAT in the glaucoma group.Results In the normal group and the post-refractive-surgery group,NCT value was(17.44±3.53) mmHg and(13.91±4.99) mmHg respectively.In the glaucoma group,GAT value was(18.90±3.74) mmHg.In all three groups,ORA values were(18.33±4.00) mmHg,(19.91±4.44) mmHg and(16.69±5.24) mmHg)respectively.ORA values were higher than NCT or GAT value in all groups.There were medium-level correlations between CCT value and NCT value in normal group and post-refractive-surgery group(r=0.466,0.472).There was a medium-level correlation between CCT values and GAT value in the glaucoma group(r=0.434).There was a low-level correlation between CCT value and ORA value in all groups(r=0.258,0.221,0.249).Conclusion Compared with the goldmann applanation tonometer and non-contact tonometer,ORA can provide more accurate IOP value and be less influenced by corneal factors.
【Key words】 Ocular response analyzer; Intraocular pressure; Goldmann applanation tonometer; Non-contact tonometer; Central corneal thickness;
- 【文献出处】 山东大学耳鼻喉眼学报 ,Journal of Otolaryngology and Ophthalmology of Shandong University , 编辑部邮箱 ,2012年04期
- 【分类号】R770.4
- 【被引频次】2
- 【下载频次】108